Review
. 2016 Apr;69(2):113-20.
doi: 10.4097/kjae.2016.69.2.113. Epub 2016 Mar 30.

Mechanisms underlying brain monitoring during anesthesia: limitations, possible improvements, and perspectives

Affiliations
Review

Mechanisms underlying brain monitoring during anesthesia: limitations, possible improvements, and perspectives

Marco Cascella. Korean J Anesthesiol. 2016 Apr.

Abstract

Currently, anesthesiologists use clinical parameters to directly measure the depth of anesthesia (DoA). This clinical standard of monitoring is often combined with brain monitoring for better assessment of the hypnotic component of anesthesia. Brain monitoring devices provide indices allowing for an immediate assessment of the impact of anesthetics on consciousness. However, questions remain regarding the mechanisms underpinning these indices of hypnosis. By briefly describing current knowledge of the brain's electrical activity during general anesthesia, as well as the operating principles of DoA monitors, the aim of this work is to simplify our understanding of the mathematical processes that allow for translation of complex patterns of brain electrical activity into dimensionless indices. This is a challenging task because mathematical concepts appear remote from clinical practice. Moreover, most DoA algorithms are proprietary algorithms and the difficulty of exploring the inner workings of mathematical models represents an obstacle to accurate simplification. The limitations of current DoA monitors - and the possibility for improvement - as well as perspectives on brain monitoring derived from recent research on corticocortical connectivity and communication are also discussed.

Keywords: Electroencephalography; Intraoperative awareness; Intraoperative neurophysiological monitoring.

PubMed Disclaimer

References

    1. Hutt A, Hudetz AG. Editorial: General anesthesia: from theory to experiments. Front Syst Neurosci. 2015;9:105. - PMC - PubMed
    1. Frost EA. Differential diagnosis of delayed awakening from general anesthesia: a review. Middle East J Anaesthesiol. 2014;22:537–548. - PubMed
    1. Pandit JJ, Andrade J, Bogod DG, Hitchman JM, Jonker WR, Lucas N, et al. 5th National Audit Project (NAP5) on accidental awareness during general anaesthesia: summary of main findings and risk factors. Br J Anaesth. 2014;113:549–559. - PubMed
    1. Chung HS. Awareness and recall during general anesthesia. Korean J Anesthesiol. 2014;66:339–345. - PMC - PubMed
    1. Musizza B, Ribaric S. Monitoring the depth of anaesthesia. Sensors (Basel) 2010;10:10896–10935. - PMC - PubMed