Does Length, Diameter, or Bone Quality Affect Primary and Secondary Stability in Self-Tapping Dental Implants?
- PMID: 27070843
- DOI: 10.1016/j.joms.2016.03.011
Does Length, Diameter, or Bone Quality Affect Primary and Secondary Stability in Self-Tapping Dental Implants?
Abstract
Purpose: Implant stability is a clinically valuable measurement of the strength of implant anchorage in the bone during placement and in the post-osseointegration period. This study aimed to determine 1) the effect of implant diameter and length and bone quality on measurements of primary and secondary stability (insertion torque [IT] and implant stability quotient [ISQ]), 2) the correlation between IT and primary and secondary ISQ, and 3) differences in ISQ in the post-osseointegration period (secondary stability) compared with immediate post-placement (primary) stability.
Patients and methods: In this longitudinal clinical study, titanium self-tapping implants were inserted in edentulous patients. The implants were grouped according to 3 independent variables: length (10 and 11.5 mm), diameter (3.75 and 4.25 mm), and bone quality (Lekholm and Zarb classification) to analyze primary and secondary implant stability (outcome variables). Statistical analyses were performed using the Student t test for paired data, 1-way analysis of variance, and the Tukey procedure for multiple pairwise comparisons.
Results: Data were collected on 88 self-tapping implants inserted in 63 partially edentulous patients. IT and implant stability were affected by diameter (3.75-mm implants, 26.5-N/cm IT and 74.0 ISQ; 4.25-mm implants, 33.8-N/cm IT and 77.0 ISQ) and bone type (type 1 + 2, 34.86-N/cm IT and 77.4 ISQ; type 3, 27.09-N/cm IT and 75.6 ISQ; type 4, 20.63-N/cm IT and 70.5 ISQ; P < .01 for all comparisons). Secondary ISQ was affected by diameter only (77.41 for 3.75- vs 75.51 for 4.25-mm implants). IT correlated with primary ISQ (R = 0.56; P < .01), although no clear correlation with secondary stability was found.
Conclusions: IT and primary ISQ in self-tapping implants differed in patients with different bone quality and implant diameter but did not differ between the 2 implant lengths compared in this study. Secondary stability was not substantially affected by any of these factors. Although IT was closely related to primary ISQ, it was unrelated to secondary ISQ. Very high primary ISQ values tended to decrease, whereas intermediate and low values tended to increase, in the transition to secondary stability.
Copyright © 2016 American Association of Oral and Maxillofacial Surgeons. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Early loading of nonsubmerged titanium implants with a chemically modified sand-blasted and acid-etched surface: 6-month results of a prospective case series study in the posterior mandible focusing on peri-implant crestal bone changes and implant stability quotient (ISQ) values.Clin Implant Dent Relat Res. 2009 Dec;11(4):338-47. doi: 10.1111/j.1708-8208.2009.00148.x. Epub 2009 Apr 23. Clin Implant Dent Relat Res. 2009. PMID: 19438966 Clinical Trial.
-
Implant Stability Quotients of Osteotome Bone Expansion and Conventional Drilling Technique for 4.1 mm Diameter Implant at Posterior Mandible.Clin Implant Dent Relat Res. 2017 Apr;19(2):253-260. doi: 10.1111/cid.12451. Epub 2016 Sep 25. Clin Implant Dent Relat Res. 2017. PMID: 27665748
-
Clinical study on the primary stability of two dental implant systems with resonance frequency analysis.Clin Oral Investig. 2007 Sep;11(3):257-65. doi: 10.1007/s00784-007-0115-2. Epub 2007 Mar 31. Clin Oral Investig. 2007. PMID: 17401588 Clinical Trial.
-
Relationship Between Primary/Mechanical and Secondary/Biological Implant Stability.Int J Oral Maxillofac Implants. 2019 Suppl;34:s7-s23. doi: 10.11607/jomi.19suppl.g1. Int J Oral Maxillofac Implants. 2019. PMID: 31116830 Review.
-
Can Osseointegration Be Achieved Without Primary Stability?Dent Clin North Am. 2019 Jul;63(3):461-473. doi: 10.1016/j.cden.2019.02.001. Epub 2019 Apr 15. Dent Clin North Am. 2019. PMID: 31097138 Review.
Cited by
-
Stability of Dental Implants and Thickness of Cortical Bone: Clinical Research and Future Perspectives. A Systematic Review.Materials (Basel). 2021 Nov 25;14(23):7183. doi: 10.3390/ma14237183. Materials (Basel). 2021. PMID: 34885335 Free PMC article. Review.
-
Immediate Loading of Single Implants in the Anterior Maxilla: A 1-Year Prospective Clinical Study on 34 Patients.Int J Dent. 2017;2017:8346496. doi: 10.1155/2017/8346496. Epub 2017 May 22. Int J Dent. 2017. PMID: 28611844 Free PMC article.
-
Repeatability of the resonance frequency analysis values in implants with a new technology.Med Oral Patol Oral Cir Bucal. 2019 Sep 1;24(5):e636-e642. doi: 10.4317/medoral.22761. Med Oral Patol Oral Cir Bucal. 2019. PMID: 31422408 Free PMC article.
-
The Interrelation between Cortical Bone Thickness and Primary and Secondary Dental Implant Stability: a Systematic Review.J Oral Maxillofac Res. 2024 Dec 31;15(4):e2. doi: 10.5037/jomr.2024.15402. eCollection 2024 Oct-Dec. J Oral Maxillofac Res. 2024. PMID: 40017687 Free PMC article. Review.
-
A Radiographic and Clinical Comparison of Immediate vs. Early Loading (4 Weeks) of Implants with a New Thermo-Chemically Treated Surface: A Randomized Clinical Trial.Int J Environ Res Public Health. 2021 Jan 29;18(3):1223. doi: 10.3390/ijerph18031223. Int J Environ Res Public Health. 2021. PMID: 33572988 Free PMC article. Clinical Trial.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources