. 2016 Sep;80(3):407-14.
doi: 10.1038/pr.2016.89. Epub 2016 Apr 13.

Parasternal intercostal electromyography: a novel tool to assess respiratory load in children

Affiliations

Parasternal intercostal electromyography: a novel tool to assess respiratory load in children

Victoria MacBean et al. Pediatr Res. 2016 Sep.

Abstract

Background: Parasternal intercostal muscle electromyography (EMGpara) represents a novel tool to assess respiratory load when volitional techniques are not possible. This study examined the application of EMGpara in healthy, wheezy, and critically ill children.

Methods: Surface EMGpara was measured during tidal breathing in 92 healthy children, 20 wheezy preschool children (with measurements repeated following bronchodilator), and 25 mechanically ventilated children during supported ventilation and on continuous positive airways pressure.

Results: EMGpara was related to age, height, and weight in the healthy group (r = -0.623, -0.625, -0.641 respectively, all P < 0.0001). An age-based equation for predicted EMGpara was developed and patient data expressed as z-scores. EMGpara was higher in wheezy children prebronchodilator than healthy controls (median interquartile range (IQR) z-score 0.53 (0.07-1.94), P = 0.0073), falling to levels not different to healthy children postbronchodilator (-0.08 (-0.50-1.00)). In the critically ill children, EMGpara was higher (P < 0.0001) than in healthy subjects during both mechanical ventilation (median (IQR) z-score 1.14 (0.33-1.93)) and continuous positive airways pressure (1.88 (0.91-3.03)).

Conclusion: EMGpara is feasible in children and infants both healthy and diseased, is raised in those with elevated respiratory load, and is responsive to clinical interventions. EMGpara represents a potential method to assess respiratory status in patients conventionally challenging to assess.

PubMed Disclaimer

References

    1. J Appl Physiol Respir Environ Exerc Physiol. 1979 May;46(5):897-904 - PubMed
    1. Exp Physiol. 2013 Jul;98(7):1190-8 - PubMed
    1. Respir Physiol. 1981 Oct;46(1):71-80 - PubMed
    1. Pediatr Pulmonol. 2008 Nov;43(11):1085-91 - PubMed
    1. Eur Respir J. 2009 Feb;33(2):289-97 - PubMed

Publication types

LinkOut - more resources