Clinical study using a new phacoemulsification system with surgical intraocular pressure control
- PMID: 27113876
- DOI: 10.1016/j.jcrs.2016.01.037
Clinical study using a new phacoemulsification system with surgical intraocular pressure control
Abstract
Purpose: To compare cumulative dissipated energy (CDE), aspiration fluid used, and aspiration time during phacoemulsification cataract extraction using 2 surgical configurations.
Setting: Two clinical sites in the United States and 1 in Spain.
Design: Prospective randomized clinical case series.
Methods: For each patient, the first eye having surgery was randomized to the active-fluidics configuration (Centurion Vision System with Active Fluidics, 0.9 mm 45-degree Intrepid Balanced tip, and 0.9 mm Intrepid Ultra infusion sleeve) or the gravity-fluidics configuration (Infiniti Vision System with gravity fluidics, 0.9 mm 45-degree Mini-Flared Kelman tip, and 0.9 mm Ultra infusion sleeve). Second-eye surgery was completed within 14 days after first-eye surgery using the alternate configuration. The CDE, aspiration fluid used, and aspiration time were compared between configurations, and adverse events were summarized.
Results: Patient demographics and cataract characteristics were similar between configurations (100 per group). The CDE was significantly lower with the active-fluidics configuration than with the gravity-fluidics configuration (mean ± standard error, 4.32 ± 0.28 percent-seconds) (P < .001). The active-fluidics configuration used significantly less aspiration fluid than the gravity-fluidics configuration (mean 46.56 ± 1.39 mL versus 52.68 ± 1.40 mL) (P < .001) and required significantly shorter aspiration time (mean 151.9 ± 4.1 seconds versus 167.6 ± 4.1 seconds) (P < .001). No serious ocular adverse events related to the study devices or device deficiencies were observed.
Conclusion: Significantly less CDE, aspiration fluid used, and aspiration time were observed with the active-fluidics configuration than with the gravity-fluidics configuration, showing improved surgical efficiency.
Financial disclosures: Drs. Solomon and Cionni are consultants to Alcon Research, Ltd., and received compensation for conduct of the study. Dr. Lorente received compensation for clinical work in the study. Mr. Fanney is an employee of Alcon Research, Ltd.
Copyright © 2016 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.
Comment in
-
Confusion about the importance of phaco tip configuration in comparing machine performance.J Cataract Refract Surg. 2016 Sep;42(9):1391. doi: 10.1016/j.jcrs.2016.07.024. J Cataract Refract Surg. 2016. PMID: 27697268 No abstract available.
Similar articles
-
Comparison of cumulative dissipated energy delivered by active-fluidic pressure control phacoemulsification system versus gravity-fluidics.Int Ophthalmol. 2018 Oct;38(5):1907-1913. doi: 10.1007/s10792-017-0674-4. Epub 2017 Aug 22. Int Ophthalmol. 2018. PMID: 28831699 Clinical Trial.
-
Effect of IOP based infusion system with and without balanced phaco tip on cumulative dissipated energy and estimated fluid usage in comparison to gravity fed infusion in torsional phacoemulsification.Eye Vis (Lond). 2017 Sep 17;4:22. doi: 10.1186/s40662-017-0087-5. eCollection 2017. Eye Vis (Lond). 2017. PMID: 28932753 Free PMC article.
-
Differences in energy expenditure for conventional and femtosecond-assisted cataract surgery using 2 different phacoemulsification systems.J Cataract Refract Surg. 2017 Jan;43(1):16-21. doi: 10.1016/j.jcrs.2016.11.037. J Cataract Refract Surg. 2017. PMID: 28317672
-
Active-fluidics versus gravity-fluidics in lens extraction: A systematic review and meta-analysis of randomized controlled trials.Eur J Ophthalmol. 2023 Jan;33(1):247-256. doi: 10.1177/11206721221107512. Epub 2022 Jun 7. Eur J Ophthalmol. 2023. PMID: 35673714
-
Application of the Active-Fluidics System in Phacoemulsification: A Review.J Clin Med. 2023 Jan 12;12(2):611. doi: 10.3390/jcm12020611. J Clin Med. 2023. PMID: 36675540 Free PMC article. Review.
Cited by
-
Comparisons of the energy efficiency and intraocular safety of two torsional phacoemulsification tips.BMC Ophthalmol. 2022 Oct 3;22(1):392. doi: 10.1186/s12886-022-02619-0. BMC Ophthalmol. 2022. PMID: 36192712 Free PMC article.
-
Comparative studies between longitudinal and torsional modes in phacoemulsification, using active fluidics technology along with the intrepid balanced tip.Indian J Ophthalmol. 2018 Oct;66(10):1417-1422. doi: 10.4103/ijo.IJO_7_18. Indian J Ophthalmol. 2018. PMID: 30249825 Free PMC article. Clinical Trial.
-
Optical coherence tomography angiography evaluation of the effects of phacoemulsification cataract surgery on macular hemodynamics in Chinese normal eyes.Int Ophthalmol. 2021 Dec;41(12):4175-4185. doi: 10.1007/s10792-021-01987-8. Epub 2021 Aug 4. Int Ophthalmol. 2021. PMID: 34350527
-
Prospective study of Centurion® versus Infiniti® phacoemulsification systems: surgical and visual outcomes.Int J Ophthalmol. 2017 Nov 18;10(11):1698-1702. doi: 10.18240/ijo.2017.11.10. eCollection 2017. Int J Ophthalmol. 2017. PMID: 29181313 Free PMC article.
-
Comparison of cumulative dissipated energy delivered by active-fluidic pressure control phacoemulsification system versus gravity-fluidics.Int Ophthalmol. 2018 Oct;38(5):1907-1913. doi: 10.1007/s10792-017-0674-4. Epub 2017 Aug 22. Int Ophthalmol. 2018. PMID: 28831699 Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical