Biometry with a new swept-source optical coherence tomography biometer: Repeatability and agreement with an optical low-coherence reflectometry device

Affiliations

Biometry with a new swept-source optical coherence tomography biometer: Repeatability and agreement with an optical low-coherence reflectometry device

Mathew Kurian et al. J Cataract Refract Surg. 2016 Apr.

Abstract

Purpose: To assess the repeatability of a new swept-source optical coherence tomography (SS-OCT) biometer (IOLMaster 700) and evaluate the agreement with an optical low-coherence reflectometry (OLCR) biometer (Lenstar 900) in cataractous eyes.

Setting: Tertiary eyecare facility.

Design: Prospective evaluation of a diagnostic test.

Methods: Eyes of patients had 3 consecutive scans acquired by the same operator. The repeatability of measurements was assessed using the within-subject standard deviation and coefficient of variation (CoV). Agreement between the mean measurements of each machine was evaluated.

Results: The study comprised 100 eyes of 100 patients. The median lens thickness and axial length (AL) measurements were significantly greater (P < .001) with SS-OCT (4.29 mm and 23.5 mm, respectively, versus 4.25 mm and 23.4 mm, respectively); anterior chamber depth (ACD) was greater with OLCR (2.72 mm versus 2.70 mm, respectively) (P < .001). The within-subject standard deviation and CoV of the AL (0.01 and 0.05, respectively) and the ACD (0.04 and 1.22, respectively) measurements showed lower variability with SS-OCT (0.05 and 0.21, respectively, for AL; 0.06 and 1.99, respectively, for ACD) than with OLCR. Good agreement was seen for anterior segment measurements between the 2 devices. The 95% limits of agreement of most parameters fell within the test-retest variability of the parameters. The SS-OCT biometer also showed better penetration in dense posterior subcapsular cataracts, measuring AL successfully in 96% of cases.

Conclusions: The new SS-OCT biometer showed good repeatability and agreement with the OLCR biometer. The SS-OCT biometer measured the AL with fewer dropouts compared with the OLCR even in dense cataracts.

Financial disclosure: None of the authors has a financial or proprietary interest in any material or method mentioned.