Meta-Analysis
. 2016 Jul;48(7):530-6.
doi: 10.3109/23744235.2016.1165350. Epub 2016 Apr 27.

Diagnostic performance of matrix-assisted laser desorption ionisation time-of-flight mass spectrometry in blood bacterial infections: a systematic review and meta-analysis

Affiliations
Meta-Analysis

Diagnostic performance of matrix-assisted laser desorption ionisation time-of-flight mass spectrometry in blood bacterial infections: a systematic review and meta-analysis

Jamie S Scott et al. Infect Dis (Lond). 2016 Jul.

Abstract

Background: Matrix-assisted laser desorption ionisation time-of-flight mass spectrometry (MALDI-TOF MS) has shown promise in decreasing time to identification of causative organisms compared to traditional methods; however, the utility of MALDI-TOF MS in a heterogeneous clinical setting is uncertain.

Objectives: To perform a systematic review on the operational performance of the Bruker MALDI-TOF MS system and evaluate published cut-off values compared to traditional blood cultures.

Methods: A comprehensive literature search was performed. Studies were included if they performed direct MALDI-TOF MS analysis of blood culture specimens in human patients with suspected bacterial infections using the Bruker Biotyper software. Sensitivities and specificities of the combined studies were estimated using a hierarchical random effects linear model (REML) incorporating cut-off scores of ≥1.7 and ≥2.0.

Results: Fifty publications were identified, with 11 studies included after final review. The estimated sensitivity utilising a cut-off of ≥2.0 from the combined studies was 74.6% (95% CI = 67.9-89.3%), with an estimated specificity of 88.0% (95% CI = 74.8-94.7%). When assessing a cut-off of ≥1.7, the combined sensitivity increases to 92.8% (95% CI = 87.4-96.0%), but the estimated specificity decreased to 81.2% (95% CI = 61.9-96.6%).

Conclusions: In this analysis, MALDI-TOF MS showed acceptable sensitivity and specificity in bacterial speciation with the current recommended cut-off point compared to blood cultures; however, lowering the cut-off point from ≥2.0 to ≥1.7 would increase the sensitivity of the test without significant detrimental effect on the specificity, which could improve clinician confidence in their results.

Keywords: Bacterial infections; MALDI-TOF MS; blood culture; matrix-assisted laser desorption ionisation time-of-flight mass spectrometry; sepsis.

PubMed Disclaimer

MeSH terms