Fluoroscopy pulse rate reduction during diagnostic and therapeutic imaging in the cardiac catheterization laboratory: An evaluation of radiation dose, procedure complications and outcomes

Affiliations

Fluoroscopy pulse rate reduction during diagnostic and therapeutic imaging in the cardiac catheterization laboratory: An evaluation of radiation dose, procedure complications and outcomes

James W Hansen et al. Catheter Cardiovasc Interv. .

Abstract

Objectives: To evaluate radiation reduction by reducing fluoroscopy pulse rate in diagnostic cardiac catheterizations and percutaneous coronary interventions (PCI) as well as outcomes at 30 days and six months.

Background: Radiation exposure to the public at large has increased dramatically over the past three decades, and the cardiac catheterization laboratory is a large contributor. Fluoroscopy pulse rate is one way to decrease radiation exposure.

Methods: Fluoroscopy pulse rate was reduced from 10 pulses/sec (p/s) to 7.5 p/s as part of an internal quality improvement project. A retrospective analysis of all cardiac catheterizations was performed, evaluating Air KERMA at the interventional reference point (Ka, r ), Air KERMA area product (PKA ), procedural complications and major adverse cardiac events at 30 days and 6 months.

Results: In diagnostic catheterization median PKA (µGy·m2 ) and Ka,r (mGy) were significantly reduced (PKA - 5,613.3 vs. 4,400, P < 0.001; Ka,r - 703.0 vs. 621.0, P = 0.041). In PCI, median PKA and Ka,r were further reduced (PKA - 13,481.6 vs. 10,648.0, P < 0.001; Ka,r - 1787.0 vs. 1,459.0, P = 0.002). There was no difference in complications, fluoroscopy time or number of stents placed. There was no difference in MACE after adjustment for number of STEMIs.

Conclusions: Reducing fluoroscopy pulse rates to 7.5 from 10 is an effective way to reduce patient radiation exposure across meaningful dose indices. A pulse rate of 7.5 p/s is safe, with no difference in complications or outcomes. A fluoroscopy pulse rate of 7.5 p/s should be given strong consideration for a new standard. © 2016 Wiley Periodicals, Inc.

Keywords: CATH, catheterization, diagnostic; CEPO, comparative effectiveness/patient centered outcomes research; PCI, percutaneous coronary intervention (PCI); RADI, radiation physics/dosimetry.