EPI Nyquist ghost and geometric distortion correction by two-frame phase labeling
- PMID: 27136196
- DOI: 10.1002/mrm.26251
EPI Nyquist ghost and geometric distortion correction by two-frame phase labeling
Abstract
Purpose: To develop a new Nyquist ghost and geometric distortion correction method in echo planar imaging (EPI) using parallel imaging.
Methods: Two frames of EPI data are acquired with normal and phase-labeled sequence. The phase label is applied by modifying the PE prephase gradient to shift the central echo by one echo spacing. GRAPPA weights are trained from both frames and used to reconstruct images from positive or negative echoes in each frame to remove Nyquist ghost. Geometric distortion is then corrected by the B0 field map generated from the phase difference between positive and negative echo images. Phantom and in vivo experiments at 7 Tesla (T) and 3T were performed to evaluate the proposed method.
Results: Nyquist ghost was greatly reduced in all images even under oblique imaging and poor eddy current conditions, yielding significant improvements over the existing reference scan and image entropy minimization based methods. Image geometries were fully restored after distortion correction. Phantom results indicated that the signal-to-noise ratio efficiency was largely preserved while fMRI results showed no apparent degradation of temporal resolution.
Conclusion: The proposed method provides robust correction of both Nyquist ghost and geometric distortion in EPI, and it is particularly suitable for dynamic EPI applications. Magn Reson Med 77:1749-1761, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
Keywords: EPI artifacts; Nyquist ghost correction; fMRI; geometric distortion correction; parallel imaging.
© 2016 International Society for Magnetic Resonance in Medicine.
Similar articles
-
Robust EPI Nyquist ghost removal by incorporating phase error correction with sensitivity encoding (PEC-SENSE).Magn Reson Med. 2018 Feb;79(2):943-951. doi: 10.1002/mrm.26710. Epub 2017 Jun 7. Magn Reson Med. 2018. PMID: 28590562
-
PEC-GRAPPA reconstruction of simultaneous multislice EPI with slice-dependent 2D Nyquist ghost correction.Magn Reson Med. 2019 Mar;81(3):1924-1934. doi: 10.1002/mrm.27546. Epub 2018 Oct 12. Magn Reson Med. 2019. PMID: 30368895
-
Dual-polarity GRAPPA for simultaneous reconstruction and ghost correction of echo planar imaging data.Magn Reson Med. 2016 Jul;76(1):32-44. doi: 10.1002/mrm.25839. Epub 2015 Jul 24. Magn Reson Med. 2016. PMID: 26208304 Free PMC article.
-
Sources of distortion in functional MRI data.Hum Brain Mapp. 1999;8(2-3):80-5. doi: 10.1002/(sici)1097-0193(1999)8:2/3<80::aid-hbm2>3.0.co;2-c. Hum Brain Mapp. 1999. PMID: 10524596 Free PMC article. Review.
-
A review of high-resolution microscopic ghost imaging with a low-dose pseudothermal light.J Microsc. 2021 Oct;284(1):3-11. doi: 10.1111/jmi.13048. Epub 2021 Aug 2. J Microsc. 2021. PMID: 34231215 Review.
Cited by
-
SENSE EPI reconstruction with 2D phase error correction and channel-wise noise removal.Magn Reson Med. 2022 Nov;88(5):2157-2166. doi: 10.1002/mrm.29349. Epub 2022 Jul 25. Magn Reson Med. 2022. PMID: 35877787 Free PMC article.
-
Ghost reduction in echo-planar imaging by joint reconstruction of images and line-to-line delays and phase errors.Magn Reson Med. 2018 Jun;79(6):3114-3121. doi: 10.1002/mrm.26967. Epub 2017 Oct 16. Magn Reson Med. 2018. PMID: 29034502 Free PMC article.
-
The impact of T1 versus EPI spatial normalization templates for fMRI data analyses.Hum Brain Mapp. 2017 Nov;38(11):5331-5342. doi: 10.1002/hbm.23737. Epub 2017 Jul 26. Hum Brain Mapp. 2017. PMID: 28745021 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources