Characteristics of Ion Activation and Collision Induced Dissociation Using Digital Ion Trap Technology
- PMID: 27150507
- DOI: 10.1007/s13361-016-1407-z
Characteristics of Ion Activation and Collision Induced Dissociation Using Digital Ion Trap Technology
Abstract
Collision induced dissociation (CID) is one of the most established techniques for tandem mass spectrometry analysis. The CID of mass selected ion could be realized by ion resonance excitation with a digital rectangular waveform. The method is simple, and highly efficient CID result could be obtained by optimizing the experimental parameters, such as digital waveform voltage, frequency, and q value. In this work, the relationship between ion trapping waveform voltage and frequency at preselected q value, the relationship between waveform frequency and the q value at certain ion trapping voltage for optimum CID efficiency were investigated. Experiment results showed that the max CID efficiency of precursor reserpine ions can be obtained at different trapping waveform voltage and frequency when q and β are different. Based on systematic experimental analysis, the optimum experimental conditions for high CID efficiency can be calculated at any selected β or q. By using digital ion trap technology, the CID process and efficient fragmentation of parent ions can be realized by simply changing the trapping waveform amplitude, frequency, and the β values in the digital ion trap mass spectrometry. The technology and method are simple. It has potential use in ion trap mass spectrometry. Graphical Abstract ᅟ.
Keywords: CID efficiency; Collision induced dissociation; Digital ion trap; Ion trap waveform frequency; Ion trap waveform voltage; q Value.
Similar articles
-
Dipolar direct current driven collision-induced dissociation in a digital ceramic-based rectilinear ion trap mass spectrometer.Anal Chem. 2013 Feb 5;85(3):1271-5. doi: 10.1021/ac3031256. Epub 2013 Jan 11. Anal Chem. 2013. PMID: 23298454
-
Resonance activation and collision-induced-dissociation of ions using rectangular wave dipolar potentials in a digital ion trap mass spectrometer.J Am Soc Mass Spectrom. 2014 Apr;25(4):556-62. doi: 10.1007/s13361-013-0804-9. Epub 2014 Jan 22. J Am Soc Mass Spectrom. 2014. PMID: 24449517
-
Synchronized Reverse Scan Collision Induced Dissociation in Digital Ion Trap Mass Spectrometer for Improving Fragment Ion Detection.Anal Chem. 2022 Dec 27;94(51):17827-17834. doi: 10.1021/acs.analchem.2c03524. Epub 2022 Dec 13. Anal Chem. 2022. PMID: 36512629
-
Ion activation methods for tandem mass spectrometry.J Mass Spectrom. 2004 Oct;39(10):1091-112. doi: 10.1002/jms.703. J Mass Spectrom. 2004. PMID: 15481084 Review.
-
Activation of large ions in FT-ICR mass spectrometry.Mass Spectrom Rev. 2005 Mar-Apr;24(2):135-67. doi: 10.1002/mas.20012. Mass Spectrom Rev. 2005. PMID: 15389858 Review.
Cited by
-
Towards Higher Sensitivity of Mass Spectrometry: A Perspective From the Mass Analyzers.Front Chem. 2021 Dec 21;9:813359. doi: 10.3389/fchem.2021.813359. eCollection 2021. Front Chem. 2021. PMID: 34993180 Free PMC article. Review.
-
Combining Micropunch Histology and Multidimensional Lipidomic Measurements for In-Depth Tissue Mapping.ACS Meas Sci Au. 2022 Feb 16;2(1):67-75. doi: 10.1021/acsmeasuresciau.1c00035. Epub 2021 Oct 28. ACS Meas Sci Au. 2022. PMID: 35647605 Free PMC article.
References
Publication types
LinkOut - more resources
Full Text Sources
Other Literature Sources