Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2016 Apr-Jun;23(2):183-6.
doi: 10.4103/0974-9233.179707.

A Comparison of Rebound to Indentation Tonometry in Supine Sedated Children with Glaucoma

Affiliations
Comparative Study

A Comparison of Rebound to Indentation Tonometry in Supine Sedated Children with Glaucoma

Dora H AlHarkan et al. Middle East Afr J Ophthalmol. 2016 Apr-Jun.

Abstract

Purpose: To compare intraocular pressure (IOP) measurements by rebound tonometry (Icare PRO; Icare Finland Oy, Helsinki, Finland) to measurements by indentation tonometry (Pneumotonometer Model 30 classic and Tono-Pen XL; Reichert, Buffalo, NY, USA) in supine sedated children with glaucoma.

Methods: Prospective comparative observational study of Saudi Arabian children with glaucoma undergoing chloral hydrate sedation for ophthalmic examination (February 2012 - February 2013). Nonglaucomatous eyes were included as controls. Eyes with corneal scars or prior corneal transplant or lamellar surgery were excluded.

Results: Fifty-two eyes (26 OD, 26 OS) of 28 children were included. Thirty-six eyes had glaucoma (32 primary congenital glaucoma, two Sturge-Weber related, and two aphakia related). Sixteen eyes did not have glaucoma (six with pediatric cataract, five normal eyes, two strabismic eyes, two eyes with simple megalocornea, and one eye had peripheral corneal laceration repair. In the glaucoma group, the mean IOP was 17.55 ± 5.97 mmHg (range, 8-31.5 mmHg) with the Icare PRO and 20 ± 6.4 mmHg (range, 8-35.5 mmHg) with the Pneumotonometer from 20.47 ± 6.81 mmHg (range, 10-43 mmHg) with the Tono-Pen XL. The Icare PRO readings were significantly lower than each of the indentation tonometers. For the control group, there was no statistically significant difference in IOP measured by the Icare PRO and the indentation tonometers.

Conclusions: In this population of supine sedated children with glaucoma, IOP measurements with the Icare PRO tend to be lower than readings from the Pneumotonometer and Tono-Pen XL.

Keywords: Chloral hydrate; Indentation tonometry; Pediatric glaucoma; Rebound tonometry.

PubMed Disclaimer

Figures

Figure 1
Figure 1
(Bland-Altman plot, Icare PRO and Pneumotonometer): The Bland-Altman plot confirms a tendency for Icare PRO values to measure lower than Pneumotonomer values
Figure 2
Figure 2
(Bland-Altman plot, Icare PRO and Tono-Pen XL): The Bland-Altman plot confirms a tendency for Icare PRO values to measure lower than Tono-Pen XL values
Figure 3
Figure 3
(Linear plot, Icare PRO and Pneumotonometer): The positive linear correlation between Icare PRO and Pneumotonometer values is shown. Spearman coefficient was 0.80, P < 0.001
Figure 4
Figure 4
(Linear plot, Icare PRO and Tono-Pen XL): The positive linear correlation between Icare PRO and Tono-Pen XL values is shown. Spearman coefficient was 0.85, P < 0.001

References

    1. Kontiola A. A new electromechanical method for measuring intraocular pressure. Doc Ophthalmol. 1996;93:265–76. - PubMed
    1. Sahin A, Basmak H, Niyaz L, Yildirim N. Reproducibility and tolerability of the ICare rebound tonometer in school children. J Glaucoma. 2007;16:185–8. - PubMed
    1. Flemmons MS, Hsiao YC, Dzau J, Asrani S, Jones S, Freedman SF. Icare rebound tonometry in children with known and suspected glaucoma. J AAPOS. 2011;15:153–7. - PubMed
    1. Gandhi NG, Prakalapakorn SG, El-Dairi MA, Jones SK, Freedman SF. Icare ONE rebound versus Goldmann applanation tonometry in children with known or suspected glaucoma. Am J Ophthalmol. 2012;154:843–9.e1. - PubMed
    1. Grigorian F, Grigorian AP, Olitsky SE. The use of the iCare tonometer reduced the need for anesthesia to measure intraocular pressure in children. J AAPOS. 2012;16:508–10. - PubMed

Publication types