. 2017 Apr;41(4):723-730.
doi: 10.1007/s00264-016-3217-0. Epub 2016 May 13.

Poor performance of Enduron polyethylene liner in total hip arthroplasty: a minimum ten-year follow up and ultra-morphological analysis of wear particles

Affiliations

Poor performance of Enduron polyethylene liner in total hip arthroplasty: a minimum ten-year follow up and ultra-morphological analysis of wear particles

Yufei Yan et al. Int Orthop. 2017 Apr.

Abstract

Purpose: The aim of the present study was to investigate the long-term outcome and the wear characteristics of two distinct types of ultra-high molecular weight polyethylene (UHMWPE) liners in total hip arthroplasty (THA).

Methods: We conducted a retrospective clinical study on patients which were treated with total hip arthroplasty using either Enduron polyethylene (Enduron PE) or Trilogy polyethylene (Trilogy PE) liners based on a minimum of ten year follow up data. Morphological analyses of wear particles from tissue samples, which were harvested during revision surgeries, were also performed.

Results: A total of 79 THAs in the Enduron group and 55 THAs in the Trilogy group were available for analysis. Kaplan-Meier survival with revision for wear-related complications as the endpoint of the Enduron PE liners was lower than that of Trilogy PE liners at ten years (93.5 % and 100 %, P = 0.03). The Enduron group had higher mean linear wear rate than that of the Trilogy group (0.20 ± 0.09 and 0.09 ± 0.03 mm/year, P < 0.01). The incidence of osteolysis for the Enduron group was higher than that of the Trilogy group (33.3 % and 12 %, P = 0.04). Under transmission electron microscopy, the Enduron group had more than 82 % of the particles less than 1.0 μm in size and more than 57 % of the particles less than 0.5 μm.

Conclusion: The long-term performance of Enduron liners was worse than that of Trilogy liners. Further clinical follow-up may be necessary in patients with Enduron PE liners in order to avoid catastrophic complications.

Keywords: Enduron polyethylene liner; Follow up; Total hip arthroplasty; Wear.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Bone Joint Surg Am. 1992 Jul;74(6):849-63 - PubMed
    1. Clin Orthop Relat Res. 2011 Aug;469(8):2262-77 - PubMed
    1. J Arthroplasty. 2005 Feb;20(2):239-43 - PubMed
    1. J Bone Joint Surg Am. 2011 Apr 20;93(8):716-22 - PubMed
    1. Acta Chir Orthop Traumatol Cech. 2010 Apr;77(2):87-92 - PubMed

Publication types

Substances

LinkOut - more resources