Effect of the Cardio First Angel™ device on CPR indices: a randomized controlled clinical trial
- PMID: 27184664
- PMCID: PMC4869179
- DOI: 10.1186/s13054-016-1296-3
Effect of the Cardio First Angel™ device on CPR indices: a randomized controlled clinical trial
Abstract
Background: A number of cardiopulmonary resuscitation (CPR) adjunct devices have been developed to improve the consistency and quality of manual chest compressions. We investigated whether a CPR feedback device would improve CPR quality and consistency, as well as patient survival.
Methods: We conducted a randomized controlled study of patients undergoing CPR for cardiac arrest in the mixed medical-surgical intensive care units of four academic teaching hospitals. Patients were randomized to receive either standard manual CPR or CPR using the Cardio First Angel™ CPR feedback device. Recorded variables included guideline adherence, CPR quality, return of spontaneous circulation (ROSC) rates, and CPR-associated morbidity.
Results: A total of 229 subjects were randomized; 149 were excluded; and 80 were included. Patient demographics were similar. Adherence to published CPR guidelines and CPR quality was significantly improved in the intervention group (p < 0.0001), as were ROSC rates (72 % vs. 35 %; p = 0.001). A significant decrease was observed in rib fractures (57 % vs. 85 %; p = 0.02), but not sternum fractures (5 % vs. 17 %; p = 0.15).
Conclusions: Use of the Cardio First Angel™ CPR feedback device improved adherence to published CPR guidelines and CPR quality, and it was associated with increased rates of ROSC. A decrease in rib but not sternum fractures was observed with device use. Further independent prospective validation is warranted to determine if these results are reproducible in other acute care settings.
Trial registration: ClinicalTrials.gov identifier: NCT02394977 . Registered on 5 Mar 2015.
Keywords: CPR outcomes; Cardiac arrest; Cardio First Angel; Cardiopulmonary resuscitation.
Figures
Comment in
-
Cardiopulmonary resuscitation quality and beyond: the need to improve real-time feedback and physiologic monitoring.Crit Care. 2016 Jun 28;20(1):182. doi: 10.1186/s13054-016-1371-9. Crit Care. 2016. PMID: 27349642 Free PMC article.
References
-
- Glatz AC, Nishisaki A, Niles DE, Hanna BD, Eilevstjonn J, Diaz LK, et al. Sternal wall pressure comparable to leaning during CPR impacts intrathoracic pressure and haemodynamics in anaesthetized children during cardiac catheterization. Resuscitation. 2013;84(12):1674–1679. doi: 10.1016/j.resuscitation.2013.07.010. - DOI - PubMed
Publication types
MeSH terms
Associated data
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical