Aims: Longer procedural time is associated with complications in radiofrequency atrial fibrillation ablation. We sought to reduce ablation time and thereby potentially reduce complications. The aim was to compare the dimensions and complications of 40 W/30 s setting to that of high-power ablations (50-80 W) for 5 s in the in vitro and in vivo models.
Methods and results: In vitro ablations-40 W/30 s were compared with 40-80 W powers for 5 s. In vivo ablations-40 W/30 s were compared with 50-80 W powers for 5 s. All in vivo ablations were performed with 10 g contact force and 30 mL/min irrigation rate. Steam pops and depth of lung lesions identified post-mortem were noted as complications. A total of 72 lesions on the non-trabeculated part of right atrium were performed in 10 Ovine. All in vitro ablations except for the 40 W/5 s setting achieved the critical lesion depth of 2 mm. For in vivo ablations, all lesions were transmural, and the lesion depths for the settings of 40 W/30 s, 50 W/5 s, 60 W/5 s, 70 W/5 s, and 80 W/5 s were 2.2 ± 0.5, 2.3 ± 0.5, 2.1 ± 0.4, 2.0 ± 0.3, and 2.3 ± 0.7 mm, respectively. The lesion depths of short-duration ablations were similar to that of the conventional ablation. Steam pops occurred in the ablation settings of 40 W/30 s and 80 W/5 s in 8 and 11% of ablations, respectively. Complications were absent in short-duration ablations of 50 and 60 W.
Conclusion: High-power, short-duration atrial ablation was as safe and effective as the conventional ablation. Compared with the conventional 40 W/30 s setting, 50 and 60 W ablation for 5 s achieved transmurality and had fewer complications.
Keywords: Atrial fibrillation ablation; High power; Radiofrequency ablation; Safety; Short duration.
Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2016. For permissions please email: journals.permissions@oup.com.
Similar articles
-
Low complication rates using high power (45-50 W) for short duration for atrial fibrillation ablations.Heart Rhythm. 2019 Feb;16(2):165-169. doi: 10.1016/j.hrthm.2018.11.031. Heart Rhythm. 2019. PMID: 30712645
-
Transmembraneous irrigation of multipolar radiofrequency ablation catheters: induction of linear lesions encircling the pulmonary vein ostium without the risk of coagulum formation?J Interv Card Electrophysiol. 2004 Jun;10(3):199-209. doi: 10.1023/B:JICE.0000026913.46734.04. J Interv Card Electrophysiol. 2004. PMID: 15133356
-
Target indices for clinical ablation in atrial fibrillation: insights from contact force, electrogram, and biophysical parameter analysis.Circ Arrhythm Electrophysiol. 2014 Feb;7(1):63-8. doi: 10.1161/CIRCEP.113.001137. Epub 2014 Jan 18. Circ Arrhythm Electrophysiol. 2014. PMID: 24443504
-
Clockwise loop of the ablation catheter in the left atrium: an easy and safe approach facilitating substrate modulation in the setting of atrial fibrillation ablation.J Cardiovasc Electrophysiol. 2008 Mar;19(3):338-41. doi: 10.1111/j.1540-8167.2007.01036.x. J Cardiovasc Electrophysiol. 2008. PMID: 18302699 Review.
-
[Complications of radiofrequency ablation: A French experience. Le Groupe de Rythmologie de la Société Francaise de Cardiologie].Arch Mal Coeur Vaiss. 1996 Dec;89(12):1599-605. Arch Mal Coeur Vaiss. 1996. PMID: 9137725 Review. French.
Cited by
-
Atrial Fibrillation Ablation: Current Practice and Future Perspectives.J Clin Med. 2023 Dec 7;12(24):7556. doi: 10.3390/jcm12247556. J Clin Med. 2023. PMID: 38137626 Free PMC article. Review.
-
Procedural Efficiency, Efficacy, and Safety of High-Power, Short-Duration Radiofrequency Ablation Delivered by STSF Catheter for Paroxysmal Atrial Fibrillation.Evid Based Complement Alternat Med. 2022 Aug 8;2022:6009275. doi: 10.1155/2022/6009275. eCollection 2022. Evid Based Complement Alternat Med. 2022. PMID: 35979002 Free PMC article.
-
The effect of different power radiofrequency ablations in treatment and postoperative pain in patients with atrial fibrillation: a retrospective study.BMC Cardiovasc Disord. 2024 Sep 9;24(1):478. doi: 10.1186/s12872-024-04147-9. BMC Cardiovasc Disord. 2024. PMID: 39251900 Free PMC article.
-
Comparative Analysis of Clinical Outcomes of High-power, Short-duration Ablation versus Low-power, Long-duration Ablation Strategy in Patients with Atrial Fibrillation: A Comprehensive Umbrella Review of Meta-analyses.J Innov Card Rhythm Manag. 2024 Aug 15;15(8):5963-5980. doi: 10.19102/icrm.2024.15083. eCollection 2024 Aug. J Innov Card Rhythm Manag. 2024. PMID: 39193534 Free PMC article. Review.
-
Long-term outcome of high-power ablation guided by ablation index in the treatment of atrial fibrillation.J Arrhythm. 2024 Apr 9;40(3):455-462. doi: 10.1002/joa3.13035. eCollection 2024 Jun. J Arrhythm. 2024. PMID: 38939788 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical