Computational fluid dynamics-based study of possibility of generating pulsatile blood flow via a continuous-flow VAD
- PMID: 27234039
- DOI: 10.1007/s11517-016-1523-8
Computational fluid dynamics-based study of possibility of generating pulsatile blood flow via a continuous-flow VAD
Abstract
Until recent years, it was almost beyond remedy to save the life of end-stage heart failure patients without considering a heart transplant. This is while the need for healthy organs has always far exceeded donations. However, the evolution of VAD technology has certainly changed the management of these patients. Today, blood pumps are designed either pulsatile flow or continuous flow, each of which has its own concerns and limitations. For instance, pulsatile pumps are mostly voluminous and hardly can be used for children. On the other hand, the flow generated by continuous-flow pumps is in contrast with pulsatile flow of the natural heart. In this project, having used computational fluid dynamics, we studied the possibility of generating pulsatile blood flow via a continuous-flow blood pump by adjusting the rotational speed of the pump with two distinct patterns (sinusoidal and trapezoidal), both of which have been proposed and set based on physiological needs and blood flow waveform of the natural heart. An important feature of this study is setting the outlet pressure of the pump similar to the physiological conditions of a patient with heart failure, and since these axial pumps are sensitive to outlet pressures, more secure and reliable results of their performance are achieved. Our results show a slight superiority of a sinusoidal pattern compared to a trapezoidal one with the potential to achieve an adequate pulsatile flow by precisely controlling the rotational speed.
Keywords: CFD; Hemolysis; Pulsatile flow; Ventricular assist device.
Similar articles
-
High-frequency operation of pulsatile ventricular assist devices: A computational study on circular and elliptically shaped pumps.Int J Artif Organs. 2019 Dec;42(12):725-734. doi: 10.1177/0391398819857442. Epub 2019 Jul 5. Int J Artif Organs. 2019. PMID: 31277562
-
Durability of left ventricular assist devices: Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) 2006 to 2011.J Thorac Cardiovasc Surg. 2013 Aug;146(2):437-41.e1. doi: 10.1016/j.jtcvs.2013.02.018. Epub 2013 Mar 13. J Thorac Cardiovasc Surg. 2013. PMID: 23490245
-
Computational characterization of flow and blood damage potential of the new maglev CH-VAD pump versus the HVAD and HeartMate II pumps.Int J Artif Organs. 2020 Oct;43(10):653-662. doi: 10.1177/0391398820903734. Epub 2020 Feb 11. Int J Artif Organs. 2020. PMID: 32043405 Free PMC article.
-
Review of recent results using computational fluid dynamics simulations in patients receiving mechanical assist devices for end-stage heart failure.Methodist Debakey Cardiovasc J. 2014 Jul-Sep;10(3):185-9. doi: 10.14797/mdcj-10-3-185. Methodist Debakey Cardiovasc J. 2014. PMID: 25574347 Free PMC article. Review.
-
Review and reflections about pulsatile ventricular assist devices from history to future: concerning safety and low haemolysis-still needed.J Artif Organs. 2020 Dec;23(4):303-314. doi: 10.1007/s10047-020-01170-3. Epub 2020 May 4. J Artif Organs. 2020. PMID: 32367384 Free PMC article. Review.
Cited by
-
Shear stress and blood trauma under constant and pulse-modulated speed CF-VAD operations: CFD analysis of the HVAD.Med Biol Eng Comput. 2019 Apr;57(4):807-818. doi: 10.1007/s11517-018-1922-0. Epub 2018 Nov 8. Med Biol Eng Comput. 2019. PMID: 30406881 Free PMC article.
-
Proposal of hemodynamically improved design of an axial flow blood pump for LVAD.Med Biol Eng Comput. 2020 Feb;58(2):401-418. doi: 10.1007/s11517-019-02097-5. Epub 2019 Dec 19. Med Biol Eng Comput. 2020. PMID: 31858420
-
Optimization of a centrifugal blood pump designed using an industrial method through experimental and numerical study.Sci Rep. 2024 Mar 28;14(1):7443. doi: 10.1038/s41598-024-57019-9. Sci Rep. 2024. PMID: 38548818 Free PMC article.
-
A Multi-Domain Simulation Study of a Pulsatile-Flow Pump Device for Heart Failure With Preserved Ejection Fraction.Front Physiol. 2022 Jan 25;13:815787. doi: 10.3389/fphys.2022.815787. eCollection 2022. Front Physiol. 2022. PMID: 35145432 Free PMC article.
-
Design and prototyping of a magnetically and hydrodynamically suspended blood pump via multiobjective optimization.Sci Rep. 2025 Jul 18;15(1):26171. doi: 10.1038/s41598-025-09828-9. Sci Rep. 2025. PMID: 40681721 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous