Value of a noise-optimized virtual monoenergetic reconstruction technique in dual-energy CT for planning of transcatheter aortic valve replacement

Affiliations

Value of a noise-optimized virtual monoenergetic reconstruction technique in dual-energy CT for planning of transcatheter aortic valve replacement

Simon S Martin et al. Eur Radiol. 2017 Feb.

Abstract

Objectives: To evaluate objective and subjective image quality of a noise-optimized virtual monoenergetic imaging (VMI+) reconstruction technique in dual-energy computed tomography (DECT) angiography prior to transcatheter aortic valve replacement (TAVR).

Methods: Datasets of 47 patients (35 men; 64.1 ± 10.9 years) who underwent DECT angiography of heart and vascular access prior to TAVR were reconstructed with standard linear blending (F_0.5), VMI+, and traditional monoenergetic (VMI) algorithms in 10-keV intervals from 40-100 keV. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of 564 arterial segments were evaluated. Subjective analysis was rated by three blinded observers using a Likert scale.

Results: Mean SNR and CNR were highest in 40 keV VMI+ series (SNR, 27.8 ± 13.0; CNR, 26.3 ± 12.7), significantly (all p < 0.001) superior to all VMI series, which showed highest values at 70 keV (SNR, 18.5 ± 7.6; CNR, 16.0 ± 7.4), as well as linearly-blended F_0.5 series (SNR, 16.8 ± 7.3; CNR, 13.6 ± 6.9). Highest subjective image quality scores were observed for 40, 50, and 60 keV VMI+ reconstructions (all p > 0.05), significantly superior to all VMI and standard linearly-blended images (all p < 0.01).

Conclusions: Low-keV VMI+ reconstructions significantly increase CNR and SNR compared to VMI and standard linear-blending image reconstruction and improve subjective image quality in preprocedural DECT angiography in the context of TAVR planning.

Key points: • VMI+ combines increased contrast with reduced image noise. • VMI+ shows substantially less image noise than traditional VMI. • 40-keV reconstructions show highest SNR/CNR of the aortic and iliofemoral access route. • Observers overall prefer 60 keV VMI+ images. • VMI+ DECT imaging helps improve image quality for TAVR planning.

Keywords: Aortic valve stenosis; Image enhancement; Multidetector-row computed tomography; Transcatheter aortic valve implantation; Transcatheter aortic valve replacement.

PubMed Disclaimer

References

    1. Eur J Radiol. 2013 Oct;82(10 ):e574-81 - PubMed
    1. Invest Radiol. 2014 Nov;49(11):735-41 - PubMed
    1. Int J Cardiovasc Imaging. 2016 Jan;32(1):137-44 - PubMed
    1. J Cardiovasc Comput Tomogr. 2015 Sep-Oct;9(5):422-7 - PubMed
    1. Eur Radiol. 2012 Dec;22(12 ):2662-9 - PubMed

MeSH terms

LinkOut - more resources