Cardiovascular implantable electronic device function and longevity at autopsy: an underestimated resource

Affiliations

Cardiovascular implantable electronic device function and longevity at autopsy: an underestimated resource

Sunil K Sinha et al. Heart Rhythm. 2016 Oct.

Abstract

Background: The feasibility and safety of postmortem cardiovascular implantable electronic device (CIED; pacemaker or defibrillator) retrieval for reuse has been shown. To date, studies indicate a low yield of reusable postmortem CIEDs (17%-30%).

Objective: The purpose of this study was to test the hypothesis that a higher rate of reusable CIEDs would be identified upon postmortem retrieval when an institutional protocol for systematic and routine acquisition, interrogation, reprogramming, and manufacturer analysis was used.

Methods: Over a 6-year period, all subjects referred for autopsy underwent concomitant CIED pulse generator retrieval and enrollment in the Johns Hopkins Post-Mortem CIED Registry. CIEDs were interrogated, reprogrammed, and submitted for manufacturer analysis.

Results: In total, 84 autopsies had CIEDs (37 pacemakers, 47 implantable cardioverter-defibrillators). CIEDs were implanted 2.84 ± 2.32 years before death, with 30% implanted <1 year before death. Overall, CIED postmortem longevity was 4.79 ± 3.41 years, with 56% demonstrating longevity ≥4 years (this group had an estimated mean longevity of 7.37 ± 2.44 years). Manufacturer analyses uncovered 2 falsely triggered elective replacement indication alerts, confirmed 5 correctly triggered elective replacement indication alerts, identified a recalled pacemaker, and verified that a defibrillator had undergone nonprogrammable hard reset.

Conclusion: When a protocol for systematic and routine postmortem CIED retrieval, interrogation, reprogramming, and analysis was used, we noted that >60% of pacemakers and >50% of defibrillators demonstrated normal functional status with projected longevities >7 years on average. Formation of a national hospital-based "CIED donor network" would facilitate larger scale charitable efforts in underserved countries.

Keywords: Autopsy; Defibrillator; Defibrillator retrieval; Pacemaker retrieval; Pacemaker reuse.