. 2016 Sep;46(3):335-43.
doi: 10.1007/s10840-016-0151-4. Epub 2016 Jun 3.

Impact of the introduction of a standardised ICD programming protocol: real-world data from a single centre

Affiliations

Impact of the introduction of a standardised ICD programming protocol: real-world data from a single centre

Nicholas Sunderland et al. J Interv Card Electrophysiol. 2016 Sep.

Abstract

Purpose: Randomised trials have shown that empiric ICD programming, using long detection times and high detection zones, reduces device therapy in ICD recipients. However, there is less data on its effectiveness in a "real-world" setting, especially secondary prevention patients. Our aim was to evaluate the introduction of a standardised programming protocol in a real-world setting of unselected ICD recipients.

Methods: We analysed 270 consecutive ICD recipients implanted in a single centre-135 implanted prior to protocol implementation (physician-led group) and 135 after (standardised group). The protocol included long arrhythmia detection times (30/40 or equivalent) and high rate detection zones (primary prevention lower treatment zone 200 bpm). Programming in the physician-led group was at the discretion of the implanter. The primary endpoint was time-to-any therapy (ATP or shocks). Secondary endpoints were time-to-inappropriate therapy and time-to-appropriate therapy. The safety endpoints were syncopal episodes, hospital admissions and death.

Results: At 12 months follow-up, 47 patients had received any ICD therapy (physician-led group, n = 31 vs. standardised group, n = 16). There was a 47 % risk reduction in any device therapy (p = 0.04) and an 86 % risk reduction in inappropriate therapy (p = 0.009) in the standardised compared to the physician-led group. There was a non-significant 30 % risk reduction in appropriate therapy (p = 0.32). Results were consistent across primary and secondary prevention patients. There were no significant differences in the rates of syncope, hospitalisation, and death.

Conclusions: In unselected patients in a real-world setting, introduction of a standardised programming protocol, using long detection times and high detection zones, significantly reduces the burden of ICD therapy without an increase in adverse outcomes.

Keywords: High detection rate; Implantable cardioverter defibrillator; Inappropriate therapy; Long detection time; Shocks.

PubMed Disclaimer

References

    1. J Am Coll Cardiol. 2008 Aug 12;52(7):541-50 - PubMed
    1. Heart Rhythm. 2015 Mar;12(3):545-53 - PubMed
    1. N Engl J Med. 2005 Jan 20;352(3):225-37 - PubMed
    1. Circulation. 2014 Jul 22;130(4):308-14 - PubMed
    1. Circulation. 2008 Sep 16;118(12):1294-303 - PubMed

MeSH terms

LinkOut - more resources