Mechanical and clinical performance of pulse-inversion tissue harmonic imaging in the superficial region
- PMID: 27278009
- DOI: 10.1007/s10396-010-0258-7
Mechanical and clinical performance of pulse-inversion tissue harmonic imaging in the superficial region
Abstract
Purpose: To compare image quality between fundamental imaging (FI; transmitted and received frequencies 8.0 and 9.0 MHz, respectively) and pulse-inversion tissue harmonic imaging (PI-THI; transmitted frequencies 4.0 and 4.5 MHz, respectively) within 40 mm of the surface of the probe, we tested axial, lateral, and contrast resolution.
Methods: The axial and lateral resolution were tested by use of a tissue-mimicking test object. For contrast resolution, milk agar blocks of different concentration were used for the comparison test.
Results: The lateral resolution was significantly improved by use of PI-THI at a depth of 20 to 40 mm from the surface of the phantom. Axial resolution was not reduced by use of PI-THI compared with FI. The value of the contrast index was higher in PI-THI than in FI. The researchers also evaluated the two modes clinically, by comparing the visibility of 170 cervical lymph nodes (identification of lymph node, echogenic hilus, parenchymal echogenicity). The results revealed significant advantages of PI-THI over FI for three features.
Conclusion: This study suggests that PI-THI might be mechanically and clinically useful in the superficial region.
Keywords: Contrast resolution; Image quality; Lymph node; Pulse inversion; Tissue harmonic image.
Similar articles
-
An Adaptive Harmonic Separation Technique for Ultrasound Harmonic Imaging.Ultrasound Med Biol. 2024 May;50(5):743-750. doi: 10.1016/j.ultrasmedbio.2024.02.001. Epub 2024 Feb 26. Ultrasound Med Biol. 2024. PMID: 38413294
-
Comparing differential tissue harmonic imaging with tissue harmonic and fundamental gray scale imaging of the liver.J Ultrasound Med. 2007 Nov;26(11):1557-63. doi: 10.7863/jum.2007.26.11.1557. J Ultrasound Med. 2007. PMID: 17957050
-
Investigation of the effect of subcutaneous fat on image quality performance of 2D conventional imaging and tissue harmonic imaging.Ultrasound Med Biol. 2005 Jul;31(7):957-64. doi: 10.1016/j.ultrasmedbio.2005.03.012. Ultrasound Med Biol. 2005. PMID: 15972201
-
A Primer on the Physical Principles of Tissue Harmonic Imaging.Radiographics. 2015 Nov-Dec;35(7):1955-64. doi: 10.1148/rg.2015140338. Radiographics. 2015. PMID: 26562232 Review.
-
Tissue harmonic imaging for standard left ventricular measurements: fundamentally flawed?Eur J Echocardiogr. 2006 Jan;7(1):9-15. doi: 10.1016/j.euje.2005.08.006. Epub 2005 Oct 10. Eur J Echocardiogr. 2006. PMID: 16219493 Review.
Cited by
-
Intraoral ultrasonography of tongue mass lesions.Dentomaxillofac Radiol. 2016;45(5):20150362. doi: 10.1259/dmfr.20150362. Epub 2016 Mar 24. Dentomaxillofac Radiol. 2016. PMID: 26915405 Free PMC article.
References
LinkOut - more resources
Full Text Sources
Research Materials
Miscellaneous