Accuracy of currently available neonatal respiratory function monitors for neonatal resuscitation
- PMID: 27279013
- PMCID: PMC4930469
- DOI: 10.1007/s00431-016-2739-1
Accuracy of currently available neonatal respiratory function monitors for neonatal resuscitation
Abstract
This study aimed to test the accuracy in volume measurements of three available respiratory function monitors (RFMs) for neonatal resuscitation and the effect of changing gas conditions. The Florian, New Life Box Neo-RSD (NLB Neo-RSD) and NICO RFM were tested on accuracy with volumes of 10 and 20 mL and on changes in volume measurements under changing gas conditions (oxygen level 21-100 % and from cold dry air (24 ± 2 °C) to heated humidified air (37 °C). Volume differences >10 % were considered clinically relevant. We found that the mean (SD) volume difference was clinically acceptable for all devices (10, 20 mL): Florian (+8.4 (1.2)%, +8.4 (0.5)%); NLB Neo-RSD (+5.8 (1.1)%, +4.3 (1.4)%); and NICO (-8.2 (0.9)%, -8.7 (0.8)%). Changing from cold dry to heated humidified air increased the volume difference using the Florian (cold dry air, heated humidified air (+5.2 (1.2)%, +12.2 (0.9)%) but not NLB Neo-RSD (+2.0(1.6)%, +3.4(2.8)%) and NICO (-2.3 % (0.8), +0.1 (0.6)%). Similarly, when using heated humidified air, increasing oxygen enlarged increased the volume difference using the Florian (oxygen 21 %, 100 %: +12.2(1.0)%, +19.8(1.1)%), but not NLB Neo-RSD (+0.2(1.9)%, +1.1(2.8)%) and NICO (-5.6(0.9)%, -3.7(0.9)%). Clinically relevant changes occurred when changing both gas conditions (Florian +25.7(1.7)%; NLB Neo-RSD +3.8(2.4)%; NICO -5.7(1.4)%).
Conclusion: The available RFMs demonstrated clinically acceptable deviations in volume measurements, except for the Florian when changing gas conditions.
What is known: •Respiratory function monitors (RFMs) are increasingly used for volume measurements during respiratory support of infants at birth. •During respiratory support at birth, gas conditions can change quickly, which can influence the volume measurements. What is new: •The available RFMs have clinically acceptable deviations when measuring the accuracy of volume measurements. •The RFM using a hot wire anemometer demonstrated clinically relevant deviations in volume measurements when changing the gas conditions. These deviations have to be taken into account when interpreting the volumes directly at birth.
Keywords: Neonatology; Respiratory function monitor; Resuscitation; Tidal volume.
Figures
Similar articles
-
Use of respiratory function monitors during simulated neonatal resuscitation.Klin Padiatr. 2011 Sep;223(5):261-6. doi: 10.1055/s-0031-1275696. Epub 2011 May 31. Klin Padiatr. 2011. PMID: 21630178
-
[Monitoring tidal volumes when using the Ventrain® emergency ventilator].Anaesthesist. 2016 Jul;65(7):514-20. doi: 10.1007/s00101-016-0161-8. Epub 2016 May 31. Anaesthesist. 2016. PMID: 27245923 German.
-
Humidified and heated air during stabilization at birth improves temperature in preterm infants.Pediatrics. 2010 Jun;125(6):e1427-32. doi: 10.1542/peds.2009-2656. Epub 2010 May 10. Pediatrics. 2010. PMID: 20457686
-
Enhanced monitoring during neonatal resuscitation.Semin Perinatol. 2019 Dec;43(8):151177. doi: 10.1053/j.semperi.2019.08.006. Epub 2019 Aug 17. Semin Perinatol. 2019. PMID: 31493856 Review.
-
Respiratory monitoring of neonatal resuscitation.Arch Dis Child Fetal Neonatal Ed. 2010 Jul;95(4):F295-303. doi: 10.1136/adc.2009.165878. Epub 2009 Sep 22. Arch Dis Child Fetal Neonatal Ed. 2010. PMID: 19776023 Review.
Cited by
-
Introducing a novel respiratory function monitor for neonatal resuscitation training.Resusc Plus. 2023 Dec 30;17:100535. doi: 10.1016/j.resplu.2023.100535. eCollection 2024 Mar. Resusc Plus. 2023. PMID: 38234876 Free PMC article.
-
The newborn delivery room of tomorrow: emerging and future technologies.Pediatr Res. 2024 Aug;96(3):586-594. doi: 10.1038/s41390-022-01988-y. Epub 2022 Mar 3. Pediatr Res. 2024. PMID: 35241791 Free PMC article. Review.
-
Ergonomic Challenges Inherent in Neonatal Resuscitation.Children (Basel). 2019 Jun 3;6(6):74. doi: 10.3390/children6060074. Children (Basel). 2019. PMID: 31163596 Free PMC article. Review.
-
Randomised study of a new inline respiratory function monitor (Juno) to improve mask seal and delivered ventilation with neonatal manikins.Arch Dis Child Fetal Neonatal Ed. 2024 Aug 16;109(5):535-541. doi: 10.1136/archdischild-2023-326256. Arch Dis Child Fetal Neonatal Ed. 2024. PMID: 38336472 Free PMC article. Clinical Trial.
-
Supraglottic airway in neonatal porcine model.Pediatr Res. 2025 Jan 21. doi: 10.1038/s41390-025-03879-4. Online ahead of print. Pediatr Res. 2025. PMID: 39837993
References
-
- Kattwinkel J, Perlman JM, Aziz K, Colby C, Fairchild K, Gallagher J, Hazinski MF, Halamek LP, Kumar P, Little G, McGowan JE, Nightengale B, Ramirez MM, Ringer S, Simon WM, Weiner GM, Wyckoff M, Zaichkin J. Part 15: Neonatal resuscitation: 2010 American Heart Association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation. 2010;122(18suppl3):s909–19. doi: 10.1161/CIRCULATIONAHA.110.971119. - DOI - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources