The Use of Micro-Computed Tomography to Determine the Accuracy of 2 Electronic Apex Locators and Anatomic Variations Affecting Their Precision
- PMID: 27291503
- DOI: 10.1016/j.joen.2016.04.024
The Use of Micro-Computed Tomography to Determine the Accuracy of 2 Electronic Apex Locators and Anatomic Variations Affecting Their Precision
Abstract
Introduction: The aim of this study was to compare the accuracy of 2 electronic apex locators (EALs), Apex ID (SybronEndo, Glendora, CA) and Root ZX (J. Morita, Tokyo, Japan), by means of micro-computed tomographic (micro-CT) imaging and to determine anatomic variations that may affect their accuracy.
Methods: The root canal length (RCL) and working length (WL) of 33 single-rooted premolars were measured using a visual method, 3-dimensional micro-CT reconstructions, and 2 different EALs. Two different measurements were recorded for each EAL: at the "APEX/0.0" mark and at the "0.5" mark. The WL was determined using 2 different methods: method 1: at the "0.5 mark" of the apex locator and method 2: subtracting 0.5 mm from the "APEX/0.0" mark. The precision of measurements was compared with those recorded by micro-CT imaging. Apical foramen (AF) position and diameter, apical constriction (AC) diameter, distance between the AC and the AF, and the presence/absence of accessory canals were recorded from the micro-CT scans, and their correlation to the accuracy of EALs was determined.
Results: There was no statistically significant difference in the RCL measurements by any of the different methods. There was a statistically significant difference in the WL recorded by micro-CT imaging compared with those by the visual method and at the "APEX/0.0 mark" - 0.5 mm (P = .031). There was no difference in the measurements acquired by any of the EALs. The "APEX/0.0 mark" - 0.5 mm was less accurate than the "0.5" mark. However, the results were not statistically significant (P > .05). The position of the AF and the AC-AF distance affected the accuracy of the RCL (P = .003) and the "0.5" mark (P = .013).
Conclusions: Root ZX and Apex ID are equally precise in determining the RCL and WL. The "0.5" mark can be used to determine the WL with high precision. Some anatomic variations may influence the accuracy of EALs.
Keywords: Apex ID; Root ZX; apical limit; electronic apex locator; micro–computed tomographic imaging; working length.
Published by Elsevier Inc.
Similar articles
-
A Micro-computed Tomographic Evaluation of the Accuracy of 3 Electronic Apex Locators in Curved Canals of Mandibular Molars.J Endod. 2018 Dec;44(12):1872-1877. doi: 10.1016/j.joen.2018.09.001. Epub 2018 Nov 1. J Endod. 2018. PMID: 30390968
-
Evaluation of the accuracy of nine electronic apex locators by using Micro-CT.Int Endod J. 2018 Feb;51(2):223-232. doi: 10.1111/iej.12814. Epub 2017 Jul 26. Int Endod J. 2018. PMID: 28675449
-
The use of micro-computed tomography to determine the accuracy of electronic working length with two apex locators.J Oral Sci. 2021 Mar 31;63(2):167-169. doi: 10.2334/josnusd.20-0466. Epub 2021 Mar 18. J Oral Sci. 2021. PMID: 33731504
-
Electronic apex locator.Dent Clin North Am. 2004 Jan;48(1):35-54. doi: 10.1016/j.cden.2003.10.005. Dent Clin North Am. 2004. PMID: 15066506 Review.
-
Integrated electronic apex locators for determining working length: a systematic review of in-vitro studies.Odontology. 2025 Jul;113(3):918-929. doi: 10.1007/s10266-024-01039-4. Epub 2024 Dec 11. Odontology. 2025. PMID: 39661212
Cited by
-
Evaluation of heated sodium hypochlorite's effect on the accuracy of contemporary electronic apex locators: an in vitro study.BMC Oral Health. 2025 Apr 24;25(1):634. doi: 10.1186/s12903-025-06003-4. BMC Oral Health. 2025. PMID: 40275207 Free PMC article.
-
Factors influencing the accuracy of electronic apex locators: A scoping review.Dent Res J (Isfahan). 2025 Jun 27;22:26. doi: 10.4103/drj.drj_634_23. eCollection 2025. Dent Res J (Isfahan). 2025. PMID: 40655901 Free PMC article. Review.
-
In Vitro Evaluation of the Accuracy of Three Electronic Apex Locators Using Different Sodium Hypochlorite Concentrations.Medicina (Kaunas). 2023 May 11;59(5):918. doi: 10.3390/medicina59050918. Medicina (Kaunas). 2023. PMID: 37241150 Free PMC article.
-
Ex Vivo Evaluation of the Accuracy of 3 Electronic Apex Locators in Different Environments: A Micro-Computed Tomography Study.Eur Endod J. 2020 Dec;5(3):226-230. doi: 10.14744/eej.2020.30633. Eur Endod J. 2020. PMID: 33353910 Free PMC article.
-
How Do Different Image Modules Impact the Accuracy of Working Length Measurements in Digital Periapical Radiography? An In Vitro Study.Diagnostics (Basel). 2025 Jan 28;15(3):305. doi: 10.3390/diagnostics15030305. Diagnostics (Basel). 2025. PMID: 39941235 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous