Case Reports
. 2014 Jan-Mar;4(1):21-3.
doi: 10.13107/jocr.2250-0685.142.

Acute Metallosis Following Total Knee Replacement - A Case Report

Affiliations
Case Reports

Acute Metallosis Following Total Knee Replacement - A Case Report

Karl C Klontz et al. J Orthop Case Rep. 2014 Jan-Mar.

Abstract

Introduction: Metallosis involving the knee joint most often results from metal-on-metal contact late in the life of a failing implant following polyethylene wear. We report a case of acute metallosis following knee arthroplasty in a previously healthy 59-year old male.

Case report: In June 2011, the patient underwent left knee arthroplasty for severe osteoarthritis with necrosis and bone edema in the medial femoral condyle and tibial plateau. Nine months later, because of persistent pain and swelling in the joint, revision arthroplasty was undertaken along with partial synovectomy. Examination revealed pristine prosthetic implants in the absence of loose fragments of bone or glue. Synovial pathology exhibited marked chronic inflammation and hyperplasia with extensive finely granular foreign material resembling metallic debris. Laboratory analysis of synovium revealed a predominance of iron, the principal component of the saw blades.

Conclusion: We hypothesize the patient experienced acute metallosis resulting from deposition of metallic fragments from three saw blades used during arthroplasty. We believe the increased density of the patient's bone that required use of multiple blades may have resulted, in part, from heavy lifting the patient partook in during the two years preceding arthroplasty.

Keywords: knee arthroplasty; metallosis; saw blades.

PubMed Disclaimer

Conflict of interest statement

Conflict of Interest: Nil

Figures

Figure 1
Figure 1
Lymphoid infiltrate and synovial lining cells in 59 year old male with acute-onset metallosis following total knee replacement (hematoxylin-eosin x 100).
Figure 2
Figure 2
Intracellular metallic debris in synovial lining cells (thin arrow) and giant cells (thick arrow) in 59 year old male with acute-onset metallosis following total knee replacement (hematoxylin-eosin x 1000).

References

    1. Romesburg JW, Wasserman PL, Schoppe CH. Metallosis and metalinduced synovitis following total knee arthroplasty: Review of radiographic and CT findings. J Radiol Case Rep. 2011;4(9):7–17. - PMC - PubMed
    1. Willis-Owen CA, Keene GC, Oakeshott RD. Early metallosis-related failure after total knee replacement: a report of 15 cases. J Bone Joint Surg Br. 2011;93(2):205–9. - PubMed
    1. Weissman BN, Scott RD, Brick GW, Corson JM. Radiographic detection of metal-induced synovitis as a complication of arthroplasty of the knee. J Bone Joint Surg. 1991;73A:1002–1007. - PubMed
    1. Mahendra G, Pandit H, Kliskey K, Murray D, Gill HS, Athanasou N. Necrotic and inflammatory changes in metal-on-metal resurfacing hip arthroplasties. Acta Orthopaedica. 2009;80(6):653–659. - PMC - PubMed
    1. Kinds MB, Marijnissen ACA, Bijlsma JWJ, Boers M, Lafeber FPJG, Welsing PMJ. Quantitative radiographic features of early knee osteoarthritis: Development over 5 years and relationship with symptoms in the CHECK Cohort. J Rheumatol. 2012 Nov 1; [epub ahead of print] - PubMed