Evaluation of the SAVI SCOUT Surgical Guidance System for Localization and Excision of Nonpalpable Breast Lesions: A Feasibility Study
- PMID: 27304083
- DOI: 10.2214/AJR.15.15962
Evaluation of the SAVI SCOUT Surgical Guidance System for Localization and Excision of Nonpalpable Breast Lesions: A Feasibility Study
Abstract
Objective: The purpose of this study is to evaluate the feasibility of the SAVI SCOUT surgical guidance system, which uses a nonradioactive infrared-activated electromagnetic wave reflector, to localize and excise nonpalpable breast lesions. We evaluated the system's use in 15 nonpalpable breast lesions in 13 patients.
Conclusion: Image-guided placement was successful for 15 of 15 (100%) reflectors. The final pathologic analysis found that lesion excision was successful, including five malignancies with negative margins. No patients required reexcision or experienced complications. SAVI SCOUT is a feasible method for breast lesion localization and excision.
Keywords: breast cancer; electromagnetic wave; infrared activated; localization.