Surgical management for retained distal embolic protection device and fractured guidewire after carotid artery stenting
- PMID: 27316621
- PMCID: PMC4911710
- DOI: 10.1093/jscr/rjw105
Surgical management for retained distal embolic protection device and fractured guidewire after carotid artery stenting
Abstract
Entrapment and fracture of carotid angioplasty and stenting hardware is a rare complication of percutaneous stenting procedures. We describe a case of a retained distal filter embolic protection device and guidewire in a 57-year-old male in Beijing, China. After unsuccessful attempts at removal via interventional methods, a second stent was deployed to secure the original hardware in situ, and the patient was discharged. He later experienced guidewire fragmentation in the carotid artery and aortic arch, with subsequent thrombus formation. We report partial removal of the guidewire and stent via carotid artery cutdown and open thoracotomy without complication. When efforts to retrieve stenting hardware are unsuccessful, it is never a suitable choice to leave them within the artery. We advocate for early surgical management of retained materials after unsuccessful carotid artery stenting. Furthermore, improved quality monitoring and assurance programs are needed to prevent such complications in the future.
Published by Oxford University Press and JSCR Publishing Ltd. All rights reserved. © The Author 2016.
Figures
References
-
- Bakoyiannis C, Economopoulos KP, Georgopoulos S, Bastounis E, Papalambros E.. Carotid endarterectomy versus carotid angioplasty with or without stenting for treatment of carotid artery stenosis: an updated meta-analysis of randomized controlled trials. Int Angiol 2010;29:205–15. - PubMed
-
- Giri J, Parikh S, Kennedy K, Weinburg I, Donaldson C, Hawkins B, et al. . Proximal versus distal embolic protection for carotid artery stenting: a national cardiovascular data registry analysis. JACC Cardiovasc Interv 2015;8:609–15. - PubMed
-
- Cremonesi A, Manetti R, Setacci F, Setacci C, Castriota F.. Protected carotid stenting: clinical advantages and complications of embolic protection devices in 442 consecutive patients. Stroke 2003;34:1936–41. - PubMed
-
- Meyers JE, Sorkin GC, Shakir HJ, Snyder KV.. Carotid cutdown for surgical retrieval of a guidewire introducer: an unusual complication after mitral valve repair. J Neurosurg 2014;121:999–1003. - PubMed