Does Magnetic Resonance Brain Scanning at 3.0 Tesla Pose a Hyperthermic Challenge to Term Neonates?
- PMID: 27318382
- DOI: 10.1016/j.jpeds.2016.05.014
Does Magnetic Resonance Brain Scanning at 3.0 Tesla Pose a Hyperthermic Challenge to Term Neonates?
Abstract
Next-generation 3-Tesla magnetic resonance (MR) scanners offer improved neonatal neuroimaging, but the greater associated radiofrequency radiation may increase the risk of hyperthermia. Safety data for neonatal 3-T MR scanning are lacking. We measured rectal temperatures continuously in 25 neonates undergoing 3-T brain MR imaging and observed no significant hyperthermic threat.
Keywords: neuroimaging; nonionizing radiation; radiofrequency radiation; specific absorption rate.
Copyright © 2016 Elsevier Inc. All rights reserved.