Comparative Study
. 2016 Nov;66(11):1109-1120.
doi: 10.1080/10962247.2016.1201022.

Comparison of real-time instruments and gravimetric method when measuring particulate matter in a residential building

Affiliations
Comparative Study

Comparison of real-time instruments and gravimetric method when measuring particulate matter in a residential building

Zuocheng Wang et al. J Air Waste Manag Assoc. 2016 Nov.

Abstract

This study used several real-time and filter-based aerosol instruments to measure PM2.5 levels in a high-rise residential green building in the Northeastern US and compared performance of those instruments. PM2.5 24-hr average concentrations were determined using a Personal Modular Impactor (PMI) with 2.5 µm cut (SKC Inc., Eighty Four, PA) and a direct reading pDR-1500 (Thermo Scientific, Franklin, MA) as well as its filter. 1-hr average PM2.5 concentrations were measured in the same apartments with an Aerotrak Optical Particle Counter (OPC) (model 8220, TSI, Inc., Shoreview, MN) and a DustTrak DRX mass monitor (model 8534, TSI, Inc., Shoreview, MN). OPC and DRX measurements were compared with concurrent 1-hr mass concentration from the pDR-1500. The pDR-1500 direct reading showed approximately 40% higher particle mass concentration compared to its own filter (n = 41), and 25% higher PM2.5 mass concentration compared to the PMI2.5 filter. The pDR-1500 direct reading and PMI2.5 in non-smoking homes (self-reported) were not significantly different (n = 10, R2 = 0.937), while the difference between measurements for smoking homes was 44% (n = 31, R2 = 0.773). Both OPC and DRX data had substantial and significant systematic and proportional biases compared with pDR-1500 readings. However, these methods were highly correlated: R2 = 0.936 for OPC versus pDR-1500 reading and R2 = 0.863 for DRX versus pDR-1500 reading. The data suggest that accuracy of aerosol mass concentrations from direct-reading instruments in indoor environments depends on the instrument, and that correction factors can be used to reduce biases of these real-time monitors in residential green buildings with similar aerosol properties.

Implications: This study used several real-time and filter-based aerosol instruments to measure PM2.5 levels in a high-rise residential green building in the northeastern United States and compared performance of those instruments. The data show that while the use of real-time monitors is convenient for measurement of airborne PM at short time scales, the accuracy of those monitors depends on a particular instrument. Bias correction factors identified in this paper could provide guidance for other studies using direct-reading instruments to measure PM concentrations.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Comparison of PM2.5 24 hr average concentrations between pDR-1500 direct reading and pDR-1500 filter-based measurements. The regression is presented by the following equation: y = β1(±SE)x + β0(±SE) where β1 and β0 are regression coefficients, and SE is the standard error of β values.
Figure 2
Figure 2
Comparison of PM2.5 24 hr average concentrations determined by PMI PM2.5 filter-based measurements and pDR-1500 filter-based measurements. The regression is presented by the following equation: y = β1(±SE)x + β0(±SE) where β1 and β0 are regression coefficients, and SE is the standard error of β values.
Figure 3
Figure 3
Comparison of PM2.5 24 hr average concentrations determined by PMI PM2.5 filter-based measurements and pDR-1500 direct reading measurements. The regression is presented by the following equation: y = β1(±SE)x + β0(±SE) where β1 and β0 are regression coefficients, and SE is the standard error of β values.
Figure 4
Figure 4
Comparison of PM2.5 24 hr average concentrations determined by PMI PM2.5 filter-based measurements and pDR-1500 direct reading measurements, with data stratified according to the self-reported smoking status in investigated apartments. The regression is presented by the following equation: y = β1(±SE)x + β0(±SE) where β1 and β0 are regression coefficients, and SE is the standard error of β values.
Figure 5
Figure 5
Comparison of PM2.5 1-hr average concentrations determined by pDR-1500 direct reading measurements and Aerotrak OPC estimate assuming particles are spherical and have a density of 1.68 g/cm3. The regression is presented by the following equation: y = β1(±SE)x + β0(±SE) where β1 and β0 are regression coefficients, and SE is the standard error of β values.
Figure 6
Figure 6
Comparison of PM2.5 1-hr average concentrations determined by pDR-1500 and DustTrak DRX direct reading measurements. The regression is presented by the following equation: y = β1(±SE)x + β0(±SE) where β1 and β0 are regression coefficients, and SE is the standard error of β values.

Similar articles

Cited by

References

    1. Abbaszadeh S, Zagreus L, Lehrer D, Huizenga C. Proceedings, Healthy Buildings 2006. Lisbon; Portugal: 2006. Occupant satisfaction with indoor environmental quality in green buildings.
    1. Adamson J, Hughes S, Azzopardi D, McAughey J, Gaca MD. Real-time assessment of cigarette smoke particle deposition in vitro. Chem Cent J. 2012;6(1):98. doi: 10.1186/1752-153X-6-98. - DOI - PMC - PubMed
    1. Ahrentzen S, Ball J, Destaillets H, Fonseca E, Fraser M, Johnson W, Patel M, Shea K. Is Green Housing Healthy Housing; A Question for Seniors. ENHR 2012 Conference (European Network for Housing Research), Workshop 19: Residential Context of Health; Lillihamer, Norway. 2012.
    1. Albanese Organization . The Solaire Residences. Marketing brochure; New York City, NY: 2004.
    1. Alderman SL, Ingebrethsen BJ. Characterization of Mainstream Cigarette Smoke Particle Size Distributions from Commercial Cigarettes Using a DMS500 Fast Particulate Spectrometer and Smoking Cycle Simulator. Aerosol Science and Technology. 2011;45(12):1409–1421. doi: 10.1080/02786826.2011.596862. - DOI