Stroke incidence in patients with cardiac implantable electronic devices remotely controlled with automatic alerts of atrial fibrillation. A sub-analysis of the HomeGuide study
- PMID: 27340918
- DOI: 10.1016/j.ijcard.2016.06.016
Stroke incidence in patients with cardiac implantable electronic devices remotely controlled with automatic alerts of atrial fibrillation. A sub-analysis of the HomeGuide study
Abstract
Background: Remote Monitoring (RM) of cardiac implantable electronic devices (CIEDs) is recommended in management of Atrial Fibrillation (AF), which is a recognized risk factor for thromboembolism. We tried to elucidate whether stroke incidence observed in a large, remotely monitored population was consistent with the CHA2DS2VASc risk profile.
Methods: Data from 1650 patients [76% male, age 72 (63-68), CHA2DS2VASc score 3.0 (2.0-4.0)] enrolled during the HomeGuide study and monitored with a daily-transmission RM system providing automatic alerts for AF, were analysed. Of those, 25% had a pacemaker and 75% an implantable cardioverter defibrillator with or without cardiac resynchronization. Estimations of the expected thromboembolic events were based on the population CHA2DS2VASc score profile used in a computer-simulated Markov model.
Results: Eight thromboembolic events were observed with a 4-year cumulative stroke rate of 0.8% (confidence interval, 0.4%-1.5%). Simulations returned from 18.7 to 17.1 expected events, depending on the AF duration assumed to trigger anticoagulation (one-sample log-rank p<0.03). During the study period, 681 (84%) AF episodes and 129 (16%) atrial tachycardias were detected in 291 patients (18%): 93% of episodes were detected remotely in 269 patients, 66% of whom had no history of AF. Medical interventions were necessary in 305 episodes, 85% of which were detected remotely. Reaction time was 1 (0-6) days for remotely-detected episodes and 33 (14-121) days for episodes detected in clinic (p<0.0001).
Conclusions: In a large CIED population followed remotely for up to 4years, the incidence of thromboembolic events was less than half the estimations based on the CHA2DS2VASc risk profile.
Keywords: Atrial fibrillation; Implantable cardioverter defibrillators; Pacemakers; Remote monitoring; Stroke; Telemedicine.
Copyright © 2016 Elsevier Ireland Ltd. All rights reserved.
Similar articles
-
Silent ischaemic brain lesions related to atrial high rate episodes in patients with cardiac implantable electronic devices.Europace. 2015 Mar;17(3):364-9. doi: 10.1093/europace/euu267. Epub 2014 Oct 21. Europace. 2015. PMID: 25336664
-
Seasonal trends in atrial fibrillation episodes and physical activity collected daily with a remote monitoring system for cardiac implantable electronic devices.Int J Cardiol. 2017 May 1;234:48-52. doi: 10.1016/j.ijcard.2017.02.074. Epub 2017 Feb 22. Int J Cardiol. 2017. PMID: 28258846
-
Home monitoring in patients with implantable cardiac devices: is there a potential reduction of stroke risk? Results from a computer model tested through monte carlo simulations.J Cardiovasc Electrophysiol. 2009 Nov;20(11):1244-51. doi: 10.1111/j.1540-8167.2009.01543.x. Epub 2009 Jul 13. J Cardiovasc Electrophysiol. 2009. PMID: 19602024
-
Subclinical atrial fibrillation and stroke: insights from continuous monitoring by implanted cardiac electronic devices.Europace. 2015 Oct;17 Suppl 2:ii40-6. doi: 10.1093/europace/euv235. Europace. 2015. PMID: 26842114 Review.
-
Methods, accuracy and clinical implications of atrial fibrillation detection by cardiac implantable electronic devices.Int J Cardiol. 2017 Jun 1;236:262-269. doi: 10.1016/j.ijcard.2016.12.189. Epub 2017 Jan 14. Int J Cardiol. 2017. PMID: 28143653 Review.
Cited by
-
2023 HRS/EHRA/APHRS/LAHRS Expert Consensus Statement on Practical Management of the Remote Device Clinic.Europace. 2023 May 19;25(5):euad123. doi: 10.1093/europace/euad123. Europace. 2023. PMID: 37208301 Free PMC article.
-
[Device-based remote monitoring : Current evidence].Herzschrittmacherther Elektrophysiol. 2017 Sep;28(3):268-278. doi: 10.1007/s00399-017-0521-3. Epub 2017 Aug 15. Herzschrittmacherther Elektrophysiol. 2017. PMID: 28812129 Review. German.
-
Remote monitoring of cardiac implantable electronic devices and disease management.Europace. 2023 Aug 2;25(9):euad233. doi: 10.1093/europace/euad233. Europace. 2023. PMID: 37622591 Free PMC article.
-
2023 HRS/EHRA/APHRS/LAHRS Expert Consensus Statement on Practical Management of the Remote Device Clinic.J Arrhythm. 2023 May 19;39(3):250-302. doi: 10.1002/joa3.12851. eCollection 2023 Jun. J Arrhythm. 2023. PMID: 37324757 Free PMC article.
-
Remote Monitoring of CIEDs-For Both Safety, Economy and Convenience?Int J Environ Res Public Health. 2021 Dec 28;19(1):312. doi: 10.3390/ijerph19010312. Int J Environ Res Public Health. 2021. PMID: 35010572 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical