A comparative planning study for lung SABR between tri-Co-60 magnetic resonance image guided radiation therapy system and volumetric modulated arc therapy
- PMID: 27401404
- DOI: 10.1016/j.radonc.2016.06.013
A comparative planning study for lung SABR between tri-Co-60 magnetic resonance image guided radiation therapy system and volumetric modulated arc therapy
Abstract
Background and purpose: To compare the plan quality of tri-(60)Co magnetic-resonance image-guided radiation therapy (IGRT) to that of volumetric modulated arc therapy (VMAT) for lung stereotactic ablative radiotherapy (SABR).
Materials and methods: A total of 22 patients with lung tumors located in the lower lobe were selected retrospectively. For each patient, VMAT plans with linac and intensity-modulated radiation therapy (IMRT) plans with the tri-(60)Co system were generated with prescription doses of 60Gy (daily dose=15Gy). For both plan types, identical CT image sets and structures were used, with the exception of planning target volumes (PTV). The PTV for VMAT was generated from the internal target volume (ITV) while the PTV for the tri-(60)Co system was generated from the gross tumor volume (GTV). Clinically relevant dose-volumetric parameters were calculated and analyzed.
Results: The average PTV volumes of tri-(60)Co plans and VMAT plans were 10.5±12.3cc vs. 27.2±23.5cc, respectively (p<0.001). The maximum and mean doses to PTVs were 64.0±2.6Gy vs. 62.5±0.9Gy (p=0.005) and 61.4±1.7Gy vs. 60.0±0.5Gy (p<0.001), respectively. The conformity and homogeneity indices were 1.89±0.38 vs. 1.01±0.40 (p<0.001) and 0.06±0.02 vs. 0.04±0.00 (p<0.001), respectively. No considerable differences for organs at risk (OARs) were observed between tri-(60)Co plans and VMAT plans. In terms of target conformity, integral dose and lung mean dose, the plan quality of tri-(60)Co plans was inferior to that of VMAT plans when the PTV volumes of tri-(60)Co plans were less than 10cc. However, all treatment plans of tri-(60)Co system were clinically acceptable.
Conclusion: For lung SABR, the quality of ITV-based VMAT plans was better than that of GTV-based tri-(60)Co plans especially when the PTV volumes of the tri-(60)Co plans were less than 10cc. If the breathing pattern of a patient is reproducible, VMAT is considered the optimal option for lung SABR, otherwise the tri-(60)Co IGRT should be considered due to the ability to monitor tumor motion during treatment.
Keywords: Lung SABR; MR-IGRT; VMAT.
Copyright © 2016 Elsevier Ireland Ltd. All rights reserved.
Similar articles
-
Quality of tri-Co-60 MR-IGRT treatment plans in comparison with VMAT treatment plans for spine SABR.Br J Radiol. 2017 Feb;90(1070):20160652. doi: 10.1259/bjr.20160652. Epub 2016 Dec 8. Br J Radiol. 2017. PMID: 27781486 Free PMC article.
-
Comparison of treatment plans between IMRT with MR-linac and VMAT for lung SABR.Radiat Oncol. 2019 Jun 13;14(1):105. doi: 10.1186/s13014-019-1314-0. Radiat Oncol. 2019. PMID: 31196120 Free PMC article.
-
Dosimetric Comparison of Real-Time MRI-Guided Tri-Cobalt-60 Versus Linear Accelerator-Based Stereotactic Body Radiation Therapy Lung Cancer Plans.Technol Cancer Res Treat. 2017 Jun;16(3):366-372. doi: 10.1177/1533034617691407. Epub 2017 Feb 7. Technol Cancer Res Treat. 2017. PMID: 28168936 Free PMC article.
-
Robust optimization in lung treatment plans accounting for geometric uncertainty.J Appl Clin Med Phys. 2018 May;19(3):19-26. doi: 10.1002/acm2.12291. Epub 2018 Mar 10. J Appl Clin Med Phys. 2018. PMID: 29524301 Free PMC article. Review.
-
LINAC based stereotactic radiosurgery for multiple brain metastases: guidance for clinical implementation.Acta Oncol. 2019 Sep;58(9):1275-1282. doi: 10.1080/0284186X.2019.1633016. Epub 2019 Jul 1. Acta Oncol. 2019. PMID: 31257960 Review.
Cited by
-
Dosimetric comparison of MR-linac-based IMRT and conventional VMAT treatment plans for prostate cancer.Radiat Oncol. 2021 Jul 21;16(1):133. doi: 10.1186/s13014-021-01858-7. Radiat Oncol. 2021. PMID: 34289868 Free PMC article.
-
Treating locally advanced lung cancer with a 1.5T MR-Linac - Effects of the magnetic field and irradiation geometry on conventionally fractionated and isotoxic dose-escalated radiotherapy.Radiother Oncol. 2017 Nov;125(2):280-285. doi: 10.1016/j.radonc.2017.09.009. Epub 2017 Oct 4. Radiother Oncol. 2017. PMID: 28987747 Free PMC article.
-
Quality of tri-Co-60 MR-IGRT treatment plans in comparison with VMAT treatment plans for spine SABR.Br J Radiol. 2017 Feb;90(1070):20160652. doi: 10.1259/bjr.20160652. Epub 2016 Dec 8. Br J Radiol. 2017. PMID: 27781486 Free PMC article.
-
Lung density change after SABR: A comparative study between tri-Co-60 magnetic resonance-guided system and linear accelerator.PLoS One. 2018 Apr 2;13(4):e0195196. doi: 10.1371/journal.pone.0195196. eCollection 2018. PLoS One. 2018. PMID: 29608606 Free PMC article.
-
MRI-guided stereotactic ablative radiation therapy of spinal bone metastases: a preliminary experience.Br J Radiol. 2020 Jan;93(1105):20190655. doi: 10.1259/bjr.20190655. Epub 2019 Nov 12. Br J Radiol. 2020. PMID: 31670569 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical