Investigation of Kryptonite™ bone cement in hybrid screw configurations of locking plate humeral midshaft fixation: A study of surrogate bone model
- PMID: 27408488
- PMCID: PMC4919364
- DOI: 10.1016/j.jor.2016.03.002
Investigation of Kryptonite™ bone cement in hybrid screw configurations of locking plate humeral midshaft fixation: A study of surrogate bone model
Abstract
Purpose: Locking compression plates (LCP) allow trauma and orthopedic surgeons to have a variety of options for utilizing locking and non-locking screw features. In this study, the hybrid constructs of mixed unicortical and bicortical screws were investigated for humerus midshaft fractures. The locking and non-locking features were also incorporated into the hybrid LCP constructs. Kryptonite™ bone cement is biocompatible with low exothermic properties and strong metal adhesion. This novel bone cement was incorporated into the non-locking screw feature to compare its mechanical effect with that of a traditional locking screw feature.
Methods: A total of 24 synthetic bones (Sawbones(®) Inc., USA) were equally divided into three groups (n = 8). The control group obtained traditional LCP fixation (JSM Medimax Inc., India) with bicortical screws, while the hybrid locking constructs employed a mix of bicortical and unicortical screws. The two bicortical screws in hybrid constructs were placed at the end holes of the LCP constructs and the unicortical screws were placed into the remaining holes. The hybrid locking (HL) group contained locking unicortical screws, whereas the hybrid non-locking group (HNK) utilized non-locking screw features incorporating the Kryptonite™ bone cement in the plate/screw bone construct. Specimens were tested by dynamic and static analysis. The eight total constructs were equally divided into two subgroups (n = 4) to conduct axial compression and torsion test individually. The low cyclic test was conducted for dynamic analysis, which included 10,000 cycles at 1 Hz frequency with a cyclic loading of 0-500 N of axial force for the axial compression test and 0-20° of angular displacement for the torsion test. The static analysis was run by a failure test with a nondestructive strain rate of 0.1 mm/s for the axial compression test and 0.5° s(-1) for the torsional test. The construct stiffness of axial compression and torsion were derived from the linear portion of the load-displacement curves. The yield strength of axial compression and torsion was determined by using offset methods. The results of stiffness and yield strength were compared by using both one-way ANOVA and Scheffe and Games-Howell post hoc tests to analyze statically significant differences among the three groups.
Results: Specimens completed 10,000 cycles in the dynamic analysis of axial and torsional cyclic tests without major deformation. To compare with the control group in static analysis, the HL and HNK groups achieved positive effect in axial stiffness, 12.3% and 10.5% greater than the control group respectively. HL obtained axial yield strength about 12% less than the control group. The HNK group was equivalent with the control group in axial yield strength. The torsional stiffness and yield strength were found similar in three groups, which indicated torsional equivalence among them. The stiffness and yield strength from axial compression and torsion were found statically non-significant among three groups.
Conclusions: Hybrid LCP constructs were initially showed to maintain equivalent axial and torsional stability with the traditional method. Kryptonite™ hosted mechanical and biological advantages for internal plate/screws (PLT) fixation.
Keywords: Internal bone fixation; Kryptonite™ bone cement; LCP; Locking vs. non-locking screw; PLT; Unicortical vs. bicortical screw.
Figures
Similar articles
-
Biomechanical evaluation of locking plate radial shaft fixation: unicortical locking fixation versus mixed bicortical and unicortical fixation in a sawbone model.J Hand Surg Am. 2007 Sep;32(7):971-5. doi: 10.1016/j.jhsa.2007.05.019. J Hand Surg Am. 2007. PMID: 17826548
-
Biomechanical Comparison of Bicortical, Unicortical, and Unicortical Far-Cortex-Abutting Screw Fixations in Plated Comminuted Midshaft Clavicle Fractures.J Hand Surg Am. 2016 Jun;41(6):703-11. doi: 10.1016/j.jhsa.2016.04.001. Epub 2016 Apr 22. J Hand Surg Am. 2016. PMID: 27113909
-
Biomechanical comparison of bicortical locking versus unicortical far-cortex-abutting locking screw-plate fixation for comminuted radial shaft fractures.J Hand Surg Am. 2014 Oct;39(10):1907-13. doi: 10.1016/j.jhsa.2014.06.141. Epub 2014 Aug 20. J Hand Surg Am. 2014. PMID: 25149898
-
Advances in Dynamization of Plate Fixation to Promote Natural Bone Healing.J Clin Med. 2024 May 14;13(10):2905. doi: 10.3390/jcm13102905. J Clin Med. 2024. PMID: 38792446 Free PMC article. Review.
-
The modified Lapidus fusion: a systematic review of biomechanical studies.EFORT Open Rev. 2023 Apr 25;8(4):162-174. doi: 10.1530/EOR-22-0069. EFORT Open Rev. 2023. PMID: 37097047 Free PMC article. Review.
Cited by
-
A biomechanical test model for evaluating osseous and osteochondral tissue adhesives.BMC Biomed Eng. 2019 May 7;1:11. doi: 10.1186/s42490-019-0011-2. eCollection 2019. BMC Biomed Eng. 2019. PMID: 32903290 Free PMC article.
-
Finite Element- and Design of Experiment-Derived Optimization of Screw Configurations and a Locking Plate for Internal Fixation System.Comput Math Methods Med. 2019 Aug 21;2019:5636528. doi: 10.1155/2019/5636528. eCollection 2019. Comput Math Methods Med. 2019. PMID: 31531124 Free PMC article.
References
-
- Denies E., Nijs S., Sermon A., Broos P. Operative treatment of humeral shaft fractures. Comparing of plating and intramedullary nailing. Acta Orthop Belg. 2010;76(6):735. - PubMed
-
- Gautier E., Perren S., Cordey J. Guidelines for the clinical application of the LCP. Injury. 2003;31:63–76. - PubMed
-
- Stoffel K., Dieter U., Stachowiak G., Gachter A., Kuster M. Biomechanical tetsing of the LCP-how can stability in locked internal fixators be controlled? Injury. 2003;34:11–19. - PubMed
-
- Patel V. Wright State University; Dayton, Ohio: 2008. Biomechanical Evaluation of Locked and Non-Locked Constructs Under Axial and Torsion Loading.
-
- Gardner M., Griffith M., Demetrakopoulous D. Hybrid locked plating of osteoporetic fractures of the humerus. J Bone Jt Surg. 2006;88(9):1962–1967. - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources