Improved standardization and potential for shortened time to results with BD Kiestra™ total laboratory automation of early urine cultures: A prospective comparison with manual processing
- PMID: 27422083
- DOI: 10.1016/j.diagmicrobio.2016.06.020
Improved standardization and potential for shortened time to results with BD Kiestra™ total laboratory automation of early urine cultures: A prospective comparison with manual processing
Abstract
We compared the results of 505 urine specimens prospectively processed by both conventional manual processing (MP) with 16-24h incubation to BD Kiestra™ Total Laboratory Automation (TLA) system with a shortened incubation of 14h: 97% of culture results were clinically concordant. TLA processing was associated with improved standardization of time of first culture reading and total incubation time.
Keywords: Automation; Kiestra™; Urine.
Crown Copyright © 2016. Published by Elsevier Inc. All rights reserved.