"Top-Out" Removal of Well-Fixed Dual-Taper Femoral Stems: Surgical Technique and Radiographic Risk Factors
- PMID: 27444847
- DOI: 10.1016/j.arth.2016.05.046
"Top-Out" Removal of Well-Fixed Dual-Taper Femoral Stems: Surgical Technique and Radiographic Risk Factors
Abstract
Background: Contemporary "dual-taper" modular femoral neck-stem designs have been associated with taper corrosion-related adverse local tissue reaction (ALTR) requiring revision surgery and stem removal. Extended trochanteric osteotomy is recognized as the workhorse procedure for revision hip surgery. The aim of our study is to describe our "top-out" stem removal surgical technique and identify preoperative radiographic risk factors associated with periprosthetic fractures when using this technique.
Methods: This is a single-center, single-surgeon, retrospective case series. Operative and clinic records were reviewed for patients with dual-taper modular femoral neck-stem junction who underwent revision surgery for taper tribocorrosion-related ALTR.
Results: Eighty-three patients (36 men and 47 women; mean age, 61.8 ± 10.3; body mass index, 30.2 ± 8.6) were revised using the top-out technique. Significant improvements in postoperative Harris hip score (P = .004), EuroQol 5-dimension questionnaire (EQ-5D; P < .001), and EQ-5D US-adjusted scores (P < .001) were observed at 19-months follow-up. Our study reports periprosthetic fracture incidence of 14% and reoperation rate of 7%. Periprosthetic fractures were positively correlated with radiographic parameters such as overhang distance (R = 0.376; P = .002) and overhang ratio (R = 0.312; P = .01) and negatively correlated with radiographic implant medial calcar prominence (R = -0.299; P = .01).
Conclusion: Removal of well-fixed femoral components can be challenging, and the burden of revision surgery for taper tribocorrosion-related ALTR of these femoral stems is likely to rise. A top-out technique with systematic preoperative planning with radiographs provides a viable, alternative surgical option to remove well-fixed femoral component while preserving the femoral bony envelope.
Keywords: adverse local tissue reaction; modular neck femoral stems; revision total hip arthroplasty; taper corrosion; top-out technique.
Copyright © 2016 Elsevier Inc. All rights reserved.
Similar articles
-
Increased risk of periprosthetic femur fractures associated with a unique cementless stem design.Clin Orthop Relat Res. 2015 Jun;473(6):2045-53. doi: 10.1007/s11999-014-4077-9. Epub 2014 Dec 12. Clin Orthop Relat Res. 2015. PMID: 25502478 Free PMC article.
-
Early Outcomes of Revision Surgery for Taper Corrosion of Dual Taper Total Hip Arthroplasty in 187 Patients.J Arthroplasty. 2016 Jul;31(7):1549-54. doi: 10.1016/j.arth.2016.01.015. Epub 2016 Jan 21. J Arthroplasty. 2016. PMID: 26895822
-
Corrosion and Fretting of a Modular Hip System: A Retrieval Analysis of 60 Rejuvenate Stems.J Arthroplasty. 2015 Aug;30(8):1470-5. doi: 10.1016/j.arth.2015.03.010. Epub 2015 Mar 14. J Arthroplasty. 2015. PMID: 25817187
-
Evaluation of the Painful Dual Taper Modular Neck Stem Total Hip Arthroplasty: Do They All Require Revision?J Arthroplasty. 2016 Jul;31(7):1385-9. doi: 10.1016/j.arth.2016.01.074. Epub 2016 Mar 21. J Arthroplasty. 2016. PMID: 27118353 Review.
-
New type of hip arthroplasty failure related to modular femoral components: breakage at the neck-stem junction.Orthop Traumatol Surg Res. 2013 Oct;99(6):741-4. doi: 10.1016/j.otsr.2013.02.010. Epub 2013 Aug 15. Orthop Traumatol Surg Res. 2013. PMID: 23953844 Review.
Cited by
-
Trunnion Failure of the Recalled Low Friction Ion Treatment Cobalt Chromium Alloy Femoral Head.J Arthroplasty. 2017 Sep;32(9):2857-2863. doi: 10.1016/j.arth.2017.03.075. Epub 2017 Apr 8. J Arthroplasty. 2017. PMID: 28478184 Free PMC article.
-
Use of Limited Femorotomy as an Alternative to Extensive Trochanteric Osteotomy for Cementless Femoral Prosthesis Revision.Arthroplast Today. 2025 Feb 25;32:101640. doi: 10.1016/j.artd.2025.101640. eCollection 2025 Apr. Arthroplast Today. 2025. PMID: 40092418 Free PMC article.
-
Dual Modular Titanium Alloy Femoral Stem Failure Mechanisms and Suggested Clinical Approaches.Materials (Basel). 2021 Jun 4;14(11):3078. doi: 10.3390/ma14113078. Materials (Basel). 2021. PMID: 34199983 Free PMC article. Review.
-
Sex- and Gender-specific Analysis in Orthopaedic Studies.Clin Orthop Relat Res. 2020 Jul;478(7):1482-1488. doi: 10.1097/CORR.0000000000001172. Clin Orthop Relat Res. 2020. PMID: 32281769 Free PMC article.
-
Outcomes following revision surgery performed for adverse reactions to metal debris in non-metal-on-metal hip arthroplasty patients: Analysis of 185 revisions from the National Joint Registry for England and Wales.Bone Joint Res. 2017 Jul;6(7):405-413. doi: 10.1302/2046-3758.67.BJR-2017-0017.R2. Bone Joint Res. 2017. PMID: 28710154 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical