Image quality comparison between single energy and dual energy CT protocols for hepatic imaging
- PMID: 27487905
- DOI: 10.1118/1.4959554
Image quality comparison between single energy and dual energy CT protocols for hepatic imaging
Abstract
Purpose: Multi-detector computed tomography (MDCT) enables volumetric scans in a single breath hold and is clinically useful for hepatic imaging. For simple tasks, conventional single energy (SE) computed tomography (CT) images acquired at the optimal tube potential are known to have better quality than dual energy (DE) blended images. However, liver imaging is complex and often requires imaging of both structures containing iodinated contrast media, where atomic number differences are the primary contrast mechanism, and other structures, where density differences are the primary contrast mechanism. Hence it is conceivable that the broad spectrum used in a dual energy acquisition may be an advantage. In this work we are interested in comparing these two imaging strategies at equal-dose and more complex settings.
Methods: We developed numerical anthropomorphic phantoms to mimic realistic clinical CT scans for medium size and large size patients. MDCT images based on the defined phantoms were simulated using various SE and DE protocols at pre- and post-contrast stages. For SE CT, images from 60 kVp through 140 with 10 kVp steps were considered; for DE CT, both 80/140 and 100/140 kVp scans were simulated and linearly blended at the optimal weights. To make a fair comparison, the mAs of each scan was adjusted to match the reference radiation dose (120 kVp, 200 mAs for medium size patients and 140 kVp, 400 mAs for large size patients). Contrast-to-noise ratio (CNR) of liver against other soft tissues was used to evaluate and compare the SE and DE protocols, and multiple pre- and post-contrasted liver-tissue pairs were used to define a composite CNR. To help validate the simulation results, we conducted a small clinical study. Eighty-five 120 kVp images and 81 blended 80/140 kVp images were collected and compared through both quantitative image quality analysis and an observer study.
Results: In the simulation study, we found that the CNR of pre-contrast SE image mostly increased with increasing kVp while for post-contrast imaging 90 kVp or lower yielded higher CNR images, depending on the differential iodine concentration of each tissue. Similar trends were seen in DE blended CNR and those from SE protocols. In the presence of differential iodine concentration (i.e., post-contrast), the CNR curves maximize at lower kVps (80-120), with the peak shifted rightward for larger patients. The combined pre- and post-contrast composite CNR study demonstrated that an optimal SE protocol has better performance than blended DE images, and the optimal tube potential for SE scan is around 90 kVp for a medium size patients and between 90 and 120 kVp for large size patients (although low kVp imaging requires high x-ray tube power to avoid photon starvation). Also, a tin filter added to the high kVp beam is not only beneficial for material decomposition but it improves the CNR of the DE blended images as well. The dose adjusted CNR of the clinical images also showed the same trend and radiologists favored the SE scans over blended DE images.
Conclusions: Our simulation showed that an optimized SE protocol produces up to 5% higher CNR for a range of clinical tasks. The clinical study also suggested 120 kVp SE scans have better image quality than blended DE images. Hence, blended DE images do not have a fundamental CNR advantage over optimized SE images.
Similar articles
-
Synthetic CT: simulating low dose single and dual energy protocols from a dual energy scan.Med Phys. 2011 Oct;38(10):5551-62. doi: 10.1118/1.3633895. Med Phys. 2011. PMID: 21992373
-
Dual-energy, standard and low-kVp contrast-enhanced CT-cholangiography: a comparative analysis of image quality and radiation exposure.Eur J Radiol. 2012 Jul;81(7):1405-12. doi: 10.1016/j.ejrad.2011.03.030. Epub 2011 Apr 2. Eur J Radiol. 2012. PMID: 21458939
-
Assessment of an advanced image-based technique to calculate virtual monoenergetic computed tomographic images from a dual-energy examination to improve contrast-to-noise ratio in examinations using iodinated contrast media.Invest Radiol. 2014 Sep;49(9):586-92. doi: 10.1097/RLI.0000000000000060. Invest Radiol. 2014. PMID: 24710203
-
Principles and Applications of Dual Energy Computed Tomography in Neuroradiology.Semin Ultrasound CT MR. 2021 Oct;42(5):418-433. doi: 10.1053/j.sult.2021.07.001. Epub 2021 Aug 1. Semin Ultrasound CT MR. 2021. PMID: 34537112 Review.
-
[A Review of Current Knowledge for X-ray Energy in CT: Practical Guide for CT Technologist].Nihon Hoshasen Gijutsu Gakkai Zasshi. 2022 May 20;78(5):449-463. doi: 10.6009/jjrt.2022-1238. Epub 2022 Apr 11. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2022. PMID: 35400711 Review. Japanese.
Cited by
-
Image Quality and Dose Comparison of Single-Energy CT (SECT) and Dual-Energy CT (DECT).Radiol Res Pract. 2020 Apr 20;2020:1403957. doi: 10.1155/2020/1403957. eCollection 2020. Radiol Res Pract. 2020. PMID: 32373363 Free PMC article. Review.
-
Single-energy versus dual-energy imaging during CT-guided biopsy using dedicated metal artifact reduction algorithm in an in vivo pig model.PLoS One. 2021 Apr 26;16(4):e0249921. doi: 10.1371/journal.pone.0249921. eCollection 2021. PLoS One. 2021. PMID: 33901200 Free PMC article.
-
Obesity-Related Pitfalls of Virtual versus True Non-Contrast Imaging-An Intraindividual Comparison in 253 Oncologic Patients.Diagnostics (Basel). 2023 Apr 26;13(9):1558. doi: 10.3390/diagnostics13091558. Diagnostics (Basel). 2023. PMID: 37174949 Free PMC article.
-
Evaluation of the best single-energy scanning in energy spectrum CT in lower extremity arteriography.Exp Ther Med. 2019 Aug;18(2):1433-1439. doi: 10.3892/etm.2019.7666. Epub 2019 Jun 11. Exp Ther Med. 2019. PMID: 31316630 Free PMC article.
-
Deep silicon photon-counting CT: A first simulation-based study for assessing perceptual benefits across diverse anatomies.Eur J Radiol. 2024 Feb;171:111279. doi: 10.1016/j.ejrad.2023.111279. Epub 2023 Dec 26. Eur J Radiol. 2024. PMID: 38194843 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical