Multicenter Study
. 2016 Dec;13(12):2306-2313.
doi: 10.1016/j.hrthm.2016.08.011. Epub 2016 Aug 5.

Catheter ablation using the third-generation cryoballoon provides an enhanced ability to assess time to pulmonary vein isolation facilitating the ablation strategy: Short- and long-term results of a multicenter study

Affiliations
Free article
Multicenter Study

Catheter ablation using the third-generation cryoballoon provides an enhanced ability to assess time to pulmonary vein isolation facilitating the ablation strategy: Short- and long-term results of a multicenter study

Arash Aryana et al. Heart Rhythm. 2016 Dec.
Free article

Abstract

Background: Limited data exist on cryoablation of atrial fibrillation (Cryo-AF) using the newly available third-generation (Arctic Front Advance-Short Tip [AFA-ST]) cryoballoon.

Objective: In this multicenter study, we evaluated the safety and efficacy of Cryo-AF using the AFA-ST vs the second-generation (Arctic Front Advance [AFA]) cryoballoon.

Methods: We examined the procedural safety and efficacy and the short- and long-term clinical outcomes associated with a first-time Cryo-AF performed in 355 consecutive patients (254/355 [72%] with paroxysmal AF), using either the AFA-ST (n = 102) or the AFA (n = 253) cryoballoon catheters.

Results: Acute isolation was achieved in 99.6% of all pulmonary veins (PVs) (AFA-ST: 100% vs AFA: 99.4%; P = .920). Time to pulmonary vein isolation was recorded in 89.2% of PVs using AFA-ST vs 60.2% using AFA (P < .001). PVs targeted using AFA-ST required fewer applications (1.6 ± 0.8 vs 1.7 ± 0.8; P = .023), whereas there were no differences in the balloon nadir temperature (AFA-ST: -47.0°C ± 7.3°C vs AFA: -47.5°C ± 7.8°C; P = .120) or thaw time (AFA-ST: 41 ± 24 seconds vs AFA: 44 ± 28 seconds; P = .056). However, AFA-ST was associated with shorter left atrial dwell time (43 ± 5 minutes vs 53 ± 16 minutes; P < .001) and procedure time (71 ± 11 minutes vs 89 ± 25 minutes; P < .001). Furthermore, Cryo-AF using AFA-ST was completed more frequently by "single-shot" PV ablation (27.4% vs 20.2%; P = .031). Persistent phrenic nerve palsy (AFA-ST: 0% vs AFA: 0.8%; P = .507) and procedure-related adverse events (AFA-ST: 1.0% vs AFA: 1.6%; P = .554) were similar, as was the freedom from recurrent atrial arrhythmias at 10 months of follow-up (AFA-ST: 81.8% vs AFA: 79.9%; P = .658).

Conclusion: Cryo-AF using the AFA-ST cryoballoon offers an enhanced ability to assess time to pulmonary vein isolation, allowing for fewer cryoapplications and shorter left atrial dwell time and procedure time. Consequently, this allowed for procedural completion more frequently using a "single-shot" PV ablation with equivalent safety and efficacy.

Keywords: Atrial fibrillation; Catheter ablation; Cryoablation; Cryoballoon; Pulmonary vein isolation.

PubMed Disclaimer

Comment in

Publication types

MeSH terms

LinkOut - more resources