Synchronized mechanical ventilation for respiratory support in newborn infants
- PMID: 27539719
- DOI: 10.1002/14651858.CD000456.pub4
Synchronized mechanical ventilation for respiratory support in newborn infants
Abstract
Background: During synchronised mechanical ventilation, positive airway pressure and spontaneous inspiration coincide. If synchronous ventilation is provoked, adequate gas exchange should be achieved at lower peak airway pressures, potentially reducing baro/volutrauma, air leak and bronchopulmonary dysplasia. Synchronous ventilation can potentially be achieved by manipulation of rate and inspiratory time during conventional ventilation and employment of patient-triggered ventilation.
Objectives: To compare the efficacy of:(i) synchronised mechanical ventilation, delivered as high-frequency positive pressure ventilation (HFPPV) or patient-triggered ventilation (assist control ventilation (ACV) and synchronous intermittent mandatory ventilation (SIMV)), with conventional ventilation or high-frequency oscillation (HFO);(ii) different types of triggered ventilation (ACV, SIMV, pressure-regulated volume control ventilation (PRVCV), SIMV with pressure support (PS) and pressure support ventilation (PSV)).
Search methods: We used the standard search strategy of the Cochrane Neonatal Review group to search the Cochrane Central Register of Controlled Trials (CENTRAL 2016, Issue 5), MEDLINE via PubMed (1966 to June 5 2016), EMBASE (1980 to June 5 2016), and CINAHL (1982 to June 5 2016). We also searched clinical trials databases, conference proceedings, and the reference lists of retrieved articles for randomised controlled trials and quasi-randomised trials.
Selection criteria: Randomised or quasi-randomised clinical trials comparing synchronised ventilation delivered as HFPPV to CMV, or ACV/SIMV to CMV or HFO in neonates. Randomised trials comparing different triggered ventilation modes (ACV, SIMV, SIMV plus PS, PRVCV and PSV) in neonates.
Data collection and analysis: Data were collected regarding clinical outcomes including mortality, air leaks (pneumothorax or pulmonary interstitial emphysema (PIE)), severe intraventricular haemorrhage (grades 3 and 4), bronchopulmonary dysplasia (BPD) (oxygen dependency beyond 28 days), moderate/severe BPD (oxygen/respiratory support dependency beyond 36 weeks' postmenstrual age (PMA) and duration of weaning/ventilation.Eight comparisons were made: (i) HFPPV versus CMV; (ii) ACV/SIMV versus CMV; (iii) SIMV or SIMV + PS versus HFO; iv) ACV versus SIMV; (v) SIMV plus PS versus SIMV; vi) SIMV versus PRVCV; vii) SIMV vs PSV; viii) ACV versus PSV. Data analysis was conducted using relative risk for categorical outcomes, mean difference for outcomes measured on a continuous scale.
Main results: Twenty-two studies are included in this review. The meta-analysis demonstrates that HFPPV compared to CMV was associated with a reduction in the risk of air leak (typical relative risk (RR) for pneumothorax was 0.69, 95% confidence interval (CI) 0.51 to 0.93). ACV/SIMV compared to CMV was associated with a shorter duration of ventilation (mean difference (MD) -38.3 hours, 95% CI -53.90 to -22.69). SIMV or SIMV + PS was associated with a greater risk of moderate/severe BPD compared to HFO (RR 1.33, 95% CI 1.07 to 1.65) and a longer duration of mechanical ventilation compared to HFO (MD 1.89 days, 95% CI 1.04 to 2.74).ACV compared to SIMV was associated with a trend to a shorter duration of weaning (MD -42.38 hours, 95% CI -94.35 to 9.60). Neither HFPPV nor triggered ventilation was associated with a significant reduction in the incidence of BPD. There was a non-significant trend towards a lower mortality rate using HFPPV versus CMV and a non-significant trend towards a higher mortality rate using triggered ventilation versus CMV. No disadvantage of HFPPV or triggered ventilation was noted regarding other outcomes.
Authors' conclusions: Compared to conventional ventilation, benefit is demonstrated for both HFPPV and triggered ventilation with regard to a reduction in air leak and a shorter duration of ventilation, respectively. In none of the trials was complex respiratory monitoring undertaken and thus it is not possible to conclude that the mechanism of producing those benefits is by provocation of synchronised ventilation. Triggered ventilation in the form of SIMV ± PS resulted in a greater risk of BPD and duration of ventilation compared to HFO. Optimisation of trigger and ventilator design with respect to respiratory diagnosis is encouraged before embarking on further trials. It is essential that newer forms of triggered ventilation are tested in randomised trials that are adequately powered to assess long-term outcomes before they are incorporated into routine clinical practice.
Update of
-
Synchronized mechanical ventilation for respiratory support in newborn infants.Cochrane Database Syst Rev. 2008 Jan 23;(1):CD000456. doi: 10.1002/14651858.CD000456.pub3. Cochrane Database Syst Rev. 2008. Update in: Cochrane Database Syst Rev. 2016 Aug 19;(8):CD000456. doi: 10.1002/14651858.CD000456.pub4. PMID: 18253979 Updated.
Similar articles
-
Synchronized mechanical ventilation for respiratory support in newborn infants.Cochrane Database Syst Rev. 2008 Jan 23;(1):CD000456. doi: 10.1002/14651858.CD000456.pub3. Cochrane Database Syst Rev. 2008. Update in: Cochrane Database Syst Rev. 2016 Aug 19;(8):CD000456. doi: 10.1002/14651858.CD000456.pub4. PMID: 18253979 Updated.
-
Synchronized mechanical ventilation for respiratory support in newborn infants.Cochrane Database Syst Rev. 2016 Sep 1;9(9):CD000456. doi: 10.1002/14651858.CD000456.pub5. Cochrane Database Syst Rev. 2016. PMID: 27581993 Free PMC article. Review.
-
Synchronized mechanical ventilation for respiratory support in newborn infants.Cochrane Database Syst Rev. 2004 Oct 18;(4):CD000456. doi: 10.1002/14651858.CD000456.pub2. Cochrane Database Syst Rev. 2004. Update in: Cochrane Database Syst Rev. 2008 Jan 23;(1):CD000456. doi: 10.1002/14651858.CD000456.pub3. PMID: 15494996 Updated.
-
Synchronized mechanical ventilation for respiratory support in newborn infants.Cochrane Database Syst Rev. 2001;(1):CD000456. doi: 10.1002/14651858.CD000456. Cochrane Database Syst Rev. 2001. Update in: Cochrane Database Syst Rev. 2004 Oct 18;(4):CD000456. doi: 10.1002/14651858.CD000456.pub2. PMID: 11279692 Updated.
-
Synchronized mechanical ventilation for respiratory support in newborn infants.Cochrane Database Syst Rev. 2000;(2):CD000456. doi: 10.1002/14651858.CD000456. Cochrane Database Syst Rev. 2000. Update in: Cochrane Database Syst Rev. 2001;(1):CD000456. doi: 10.1002/14651858.CD000456. PMID: 10796368 Updated.
Cited by
-
[Safety of two ventilator weaning strategies after high-frequency oscillatory ventilation in preterm infants with respiratory distress syndrome: a prospective randomized controlled trial].Zhongguo Dang Dai Er Ke Za Zhi. 2021 Jan;23(1):18-24. doi: 10.7499/j.issn.1008-8830.2010115. Zhongguo Dang Dai Er Ke Za Zhi. 2021. PMID: 33476532 Free PMC article. Clinical Trial. Chinese.
-
Therapeutic effects and outcomes of rescue high-frequency oscillatory ventilation for premature infants with severe refractory respiratory failure.Sci Rep. 2021 Apr 19;11(1):8471. doi: 10.1038/s41598-021-88231-6. Sci Rep. 2021. PMID: 33875758 Free PMC article. Clinical Trial.
-
Environmental factors contribute to the success of extubation in newborns: A multicenter cross-sectional study.Braz J Phys Ther. 2025 Aug 12;29(6):101255. doi: 10.1016/j.bjpt.2025.101255. Online ahead of print. Braz J Phys Ther. 2025. PMID: 40803259 Free PMC article.
-
[Research advances in the methods for weaning from high-frequency oscillatory ventilation in neonates].Zhongguo Dang Dai Er Ke Za Zhi. 2019 Dec;21(12):1234-1238. doi: 10.7499/j.issn.1008-8830.2019.12.016. Zhongguo Dang Dai Er Ke Za Zhi. 2019. PMID: 31874666 Free PMC article. Review. Chinese.
-
Evaluation of lung homogeneity in neonates and small infants during general anaesthesia using electrical impedance tomography: a prospective observational study.BJA Open. 2024 Sep 21;12:100344. doi: 10.1016/j.bjao.2024.100344. eCollection 2024 Dec. BJA Open. 2024. PMID: 39364288 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources