Comparative Study
. 2017 Jan;103(1):139-144.
doi: 10.1016/j.athoracsur.2016.05.089. Epub 2016 Aug 17.

Comparison of the Performance of a Sutureless Bioprosthesis With Two Pericardial Stented Valves on Small Annuli: An In Vitro Study

Affiliations
Comparative Study

Comparison of the Performance of a Sutureless Bioprosthesis With Two Pericardial Stented Valves on Small Annuli: An In Vitro Study

Giordano Tasca et al. Ann Thorac Surg. 2017 Jan.

Abstract

Background: Aortic valve replacement has evolved recently with the development of the sutureless bioprosthesis. One such valve is the Perceval bioprosthesis, which is built by mounting leaflets of bovine pericardium to a thin stent; this approach has the potential to provide an excellent fluid dynamic performance. We undertook an in vitro study to compare the hydrodynamic performance of the sutureless bioprosthesis with two standard pericardial stented bioprostheses (Crown and Magna).

Methods: Tests were conducted using a mock loop, testing on two sizes of the three prostheses. The prosthesis sizes were chosen to house the valves in porcine aortic roots with a native annulus diameter of 19 mm (n = 6) or 21 mm (n = 6). The stroke volume ranged from 25 mL to 105 mL at a simulated heart rate of 70 beats per minute.

Results: Mean pressure drop and energy loss rose with increasing stroke volume in all of the valves tested (p < 0.001), with the sutureless valve showing the lowest values for both variables (p < 0.001). Effective orifice area values were stable across the stroke volume intervals and were larger in the sutureless valves (p < 0.001).

Conclusions: All of the valves tested provided good fluid dynamic performances. The sutureless bioprosthesis provided the best performance with the least hindrance to flow behavior. From the hydrodynamic perspective, the sutureless prosthesis may present an advance in the evolution of bioprostheses, ensuring low gradients and potential for low incidence of patient-prosthesis mismatch even in small annuli.