Multicenter Study
. 2016 Dec;102(6):1956-1965.
doi: 10.1016/j.athoracsur.2016.05.080. Epub 2016 Aug 17.

Pericardial Stentless Valve for Aortic Valve Replacement: Long-Term Results

Affiliations
Multicenter Study

Pericardial Stentless Valve for Aortic Valve Replacement: Long-Term Results

Alberto Repossini et al. Ann Thorac Surg. 2016 Dec.

Abstract

Background: The Freedom Solo (FS) bovine pericardial valve (Sorin Group, Milan, Italy) is a stentless bioprosthesis that was introduced in 2004 and approved by the United States Food and Drug Administration in 2014. No long-term follow-up series are available to date. We report the multicenter experience of 4 European institutions that began implanting FS extensively from its introduction, providing the largest series with long-term follow-up.

Methods: From 2004 to 2009, 565 patients (242 women [42.8%]; mean age, 74.6 ± 8.3 years) underwent isolated (n = 350) or combined (n = 215) aortic valve replacement with the FS. Mean follow-up, including clinical and strict echocardiographic evaluation, was 6.9 ± 3.7 years (maximum, 11.8 years; cumulative 2,965 patient-years). Primary end point was freedom from structural valve deterioration (SVD), and secondary end points were freedom from reoperation and overall survival.

Results: Mean logistic European System for Cardiac Operative Risk Evaluation I was 10.3% ± 6.7%. Overall 30-day mortality was 3.7%, and no deaths were valve related. There was no severe prostheses-patient mismatch, and moderate prostheses-patient mismatch occurred only in 1 patient (0.17%). Twenty-eight patients (5.2%) underwent reoperation (20 surgical replacements, 8 transcatheter aortic valve-in-valve replacements) due to endocarditis in 9, blunt trauma in 1, and SVD in 18. SVD was reported in 5 other patients alive at time of censoring. Freedom from SVD and reoperation was 90.8% (95% confidence interval, 89.1% to 92.5%) and 87.3% (95% confidence interval, 85.6% to 89.0%), respectively, at 10 years of follow-up, and the overall actuarial survival was 56.4% (95% confidence interval, 53.3% to 59.5%).

Conclusions: The FS valve provided excellent long-term durability and hemodynamic performance in this large, multicenter European experience. Moreover, the FS, given the low rate of SVD, along with a simple implantability, proved to be a reliable bioprosthesis in the aortic position as a valid alternative to stented bioprostheses.

PubMed Disclaimer

MeSH terms

Substances

LinkOut - more resources