Do Modern Hearing Aids Meet ANSI Standards?
- PMID: 27564440
- DOI: 10.3766/jaaa.15027
Do Modern Hearing Aids Meet ANSI Standards?
Abstract
Background: The American National Standards Institute (ANSI) provides standards used to govern standardization of all hearing aids. If hearing aids do not meet specifications, there are potential negative implications for hearing aid users, professionals, and the industry. Recent literature has not investigated the proportion of new hearing aids in compliance with the ANSI specifications for quality control standards when they arrive in the clinic before dispensing.
Purpose: The aims of this study were to determine the percentage of new hearing aids compliant with the relevant ANSI standard and to report trends in electroacoustic analysis data.
Research design: New hearing aids were evaluated for quality control via the ANSI S3.22-2009 standard. In addition, quality control of directional processing was also assessed.
Study sample: Seventy-three behind-the-ear hearing aids from four major manufacturers, that were purchased for clinical patients were evaluated before dispensing.
Data collection and analysis: Audioscan Verifit (version 3.1) hearing instrument fitting system was used to complete electroacoustic analysis and directional processing evaluation of the hearing aids. Frye's Fonix 8000 test box system (Fonix 8000) was also used to cross-check equivalent input noise (EIN) measurements. These measurements were then analyzed for trends across brands and specifications.
Results: All of the hearing aids evaluated were found to be out of specification for at least one measure. EIN and attack and release times were the measures most frequently out of specification. EIN was found to be affected by test box isolation for two of the four brands tested. Systematic discrepancies accounted for ∼93% of the noncompliance issues, while unsystematic quality control issues accounted for the remaining 7%.
Conclusions: The high number of systematic discrepancies between the data collected and the specifications published by the manufacturers suggests there are clear issues related to the specific protocols used for quality control testing. These issues present a significant barrier for hearing aid dispensers when attempting to accurately determine if a hearing aid is functioning appropriately. The significant number of unsystematic discrepancies supports the continued importance of quality control measures of new and repaired hearing aids to ensure that the device is functioning properly before it is dispensed and to avoid future negative implications of fitting a faulty device.
American Academy of Audiology.
Similar articles
-
Performance of new hearing aids using the ANSI S3.22-1976 standard.J Speech Hear Disord. 1982 Nov;47(4):376-82. doi: 10.1044/jshd.4704.376. J Speech Hear Disord. 1982. PMID: 7186579
-
Telecoil-mode hearing aid compatibility performance requirements for wireless and cordless handsets: magnetic signal levels.J Am Acad Audiol. 2011 Sep;22(8):515-27. doi: 10.3766/jaaa.22.8.4. J Am Acad Audiol. 2011. PMID: 22031676 Clinical Trial.
-
High-frequency audibility: the effects of audiometric configuration, stimulus type, and device.J Am Acad Audiol. 2015 Feb;26(2):128-37. doi: 10.3766/jaaa.26.2.3. J Am Acad Audiol. 2015. PMID: 25690773 Free PMC article.
-
Open Versus Closed Hearing-Aid Fittings: A Literature Review of Both Fitting Approaches.Trends Hear. 2016 Feb 15;20:2331216516631741. doi: 10.1177/2331216516631741. Trends Hear. 2016. PMID: 26879562 Free PMC article. Review.
-
Technical and audiological factors in the implementation and use of digital signal processing hearing aids.Scand Audiol. 1997;26(4):223-9. doi: 10.3109/01050399709048011. Scand Audiol. 1997. PMID: 9428031 Review.
Cited by
-
Applications of direct-to-consumer hearing devices for adults with hearing loss: a review.Clin Interv Aging. 2017 May 18;12:859-871. doi: 10.2147/CIA.S135390. eCollection 2017. Clin Interv Aging. 2017. PMID: 28553093 Free PMC article.
-
LoCHAid: An ultra-low-cost hearing aid for age-related hearing loss.PLoS One. 2020 Sep 23;15(9):e0238922. doi: 10.1371/journal.pone.0238922. eCollection 2020. PLoS One. 2020. PMID: 32966301 Free PMC article.
-
Verification and Validation: Just the Standards.Semin Hear. 2022 Jul 26;43(2):85-93. doi: 10.1055/s-0042-1749208. eCollection 2022 May. Semin Hear. 2022. PMID: 35903074 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical