. 2016 Oct;254(10):2017-2023.
doi: 10.1007/s00417-016-3453-4. Epub 2016 Aug 27.

Epiretinal membrane as a source of errors during the measurement of peripapillary nerve fibre thickness using spectral-domain optical coherence tomography (SD-OCT)

Affiliations

Epiretinal membrane as a source of errors during the measurement of peripapillary nerve fibre thickness using spectral-domain optical coherence tomography (SD-OCT)

Florian Rüfer et al. Graefes Arch Clin Exp Ophthalmol. 2016 Oct.

Abstract

Purpose: We aimed to examine the extent to which measurement errors in the determination of retinal nerve fibre layer (RNFL) using spectral-domain optical coherence tomography (SD-OCT) occur in cases of epiretinal membrane and whether systematic deviations are found in the values obtained.

Methods: A macular scan and a circumpapillary scan were performed on 97 eyes of 97 patients using SD-OCT. Group 1 comprised 53 patients with epiretinal membrane at an age of 70 ± 4.8 years (median ± average absolute deviation). Group 2 consisted of 44 patients without any macular pathologies (median age 70 ± 5.8 years). Differences in the thickness of the RNFL and segmentation errors in the detection of the RNFL were recorded quantitatively in both groups and checked for statistical significance using non-parametric tests.

Results: The median central retinal thickness in Group 1 was 357 ± 79 μm (median ± average absolute deviation), and in Group 2 it was 270 ± 11 μm (p < 0.001). The result of the quadrant-by-quadrant measurement of the average RNFL in Group 1 and Group 2, respectively, was: temporal 88 ± 17 and 73 ± 9 μm, inferior 121 ± 17 and 118 ± 15 μm, nasal 87 ± 15 and 89 ± 14 μm and superior 115 ± 15 and 114 ± 9 μm. Temporally, the difference was statistically significant (p < 0.001). Segmentation errors of the RNFL were found in 19 of 53 eyes (35.8 %) in Group 1 and in no eyes (p < 0.001) in Group 2.

Conclusions: In eyes with epiretinal membrane, measuring errors in the SD-OCT occur significantly more frequently than in eyes without any retinal pathologies. If epiretinal membrane and glaucoma are present simultaneously, the results of the automated RNFL measurement using SD-OCT should be critically scrutinised, even if no papillary changes are visible clinically.

Keywords: Epiretinal membrane; Measuring error; Optical coherence tomography; Peripapillary nerve fibre thickness.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Arch Ophthalmol. 1991 Jan;109(1):77-83 - PubMed
    1. Klin Monbl Augenheilkd. 2006 Aug;223(8):687-90 - PubMed
    1. Eye (Lond). 2010 Sep;24(9):1498-502 - PubMed
    1. Curr Opin Ophthalmol. 2013 Mar;24(2):150-61 - PubMed
    1. Ophthalmic Surg Lasers Imaging. 2010 Nov-Dec;41 Suppl:S15-27 - PubMed

LinkOut - more resources