A novel color-LED corneal topographer to assess astigmatism in pseudophakic eyes
- PMID: 27574391
- PMCID: PMC4990375
- DOI: 10.2147/OPTH.S113027
A novel color-LED corneal topographer to assess astigmatism in pseudophakic eyes
Abstract
Purpose: To assess the accuracy of corneal astigmatism evaluation measured by four techniques, Orbscan IIz(®), Lenstar LS900(®), Cassini(®), and Total Cassini (anterior + posterior surface), in pseudophakic eyes.
Patients and methods: A total of 30 patients (46 eyes) who had undergone cataract surgery with the implantation of a monofocal intraocular lens (AcrySof IQ) were assessed after surgery. For each eye, subjective assessment of astigmatism and its axis was performed. Minimum, maximum, and mean keratometry and astigmatism and its axis were evaluated using the four measurement techniques. All measurements were compared with the subjective measurements. Agreement between each measurement technique and subjective assessment was evaluated using Bland-Altman plots. Linear regressions were performed and compared.
Results: Linear regression analysis of astigmatism axis showed very high R (2) for all models, with Total Cassini showing the least difference to the unit slope (0.052) and the least difference to a null constant (3.790), although not statistically different from the other models. Regarding astigmatism value, the Cassini and Total Cassini models were similar and statistically better than the Lenstar model. Cassini and Total Cassini showed better J0 compared with Orbscan.
Conclusion: On linear regression models, Cassini and Total Cassini showed the best performance regarding astigmatism value. Cassini and Total Cassini also showed the least J0 deviation from the Cartesian origin compared with Orbscan, which had the lowest performance. Total corneal measurement with the color LED topographer seems to be a better technique for astigmatism assessment.
Keywords: astigmatism; keratometry; topography.
Figures





Similar articles
-
Comparability and repeatability of different methods of corneal astigmatism assessment.Clin Ophthalmol. 2017 Dec 20;12:29-34. doi: 10.2147/OPTH.S146730. eCollection 2018. Clin Ophthalmol. 2017. PMID: 29339918 Free PMC article.
-
Comparison of keratometry measurements using the Pentacam HR, the Orbscan IIz, and the TMS-4 topographer.Ophthalmic Physiol Opt. 2012 Nov;32(6):539-46. doi: 10.1111/j.1475-1313.2012.00942.x. Ophthalmic Physiol Opt. 2012. PMID: 23057566
-
Comparison of keratometric measurements between color light-emitting diode topography and Scheimpflug camera.BMC Ophthalmol. 2019 Apr 26;19(1):98. doi: 10.1186/s12886-019-1106-1. BMC Ophthalmol. 2019. PMID: 31027491 Free PMC article.
-
Repeatability and comparability of corneal power and corneal astigmatism obtained from a point-source color light-emitting diode topographer, a Placido-based corneal topographer, and a low-coherence reflectometer.J Cataract Refract Surg. 2015 Oct;41(10):2242-50. doi: 10.1016/j.jcrs.2015.11.003. J Cataract Refract Surg. 2015. PMID: 26703301
-
Accuracy of Total Corneal Astigmatism Measurements With a Scheimpflug Imager and a Color Light-Emitting Diode Corneal Topographer.Am J Ophthalmol. 2016 Jul;167:72-8. doi: 10.1016/j.ajo.2016.04.011. Epub 2016 Apr 27. Am J Ophthalmol. 2016. PMID: 27131773
Cited by
-
[Early diagnosis of keratoconus].Ophthalmologe. 2022 Feb;119(2):209-218. doi: 10.1007/s00347-021-01460-y. Epub 2021 Jul 23. Ophthalmologe. 2022. PMID: 34297190 German.
-
How Can We Improve Toric Intraocular Lens Calculation Methods? Current Insights.Clin Ophthalmol. 2020 Jul 6;14:1899-1908. doi: 10.2147/OPTH.S238686. eCollection 2020. Clin Ophthalmol. 2020. PMID: 32753834 Free PMC article. Review.
-
A review of corneal imaging methods for the early diagnosis of pre-clinical Keratoconus.J Optom. 2020 Oct-Dec;13(4):269-275. doi: 10.1016/j.optom.2019.11.001. Epub 2020 Jan 6. J Optom. 2020. PMID: 31917136 Free PMC article. Review.
-
Current Developments in Corneal Topography and Tomography.Diagnostics (Basel). 2021 Aug 13;11(8):1466. doi: 10.3390/diagnostics11081466. Diagnostics (Basel). 2021. PMID: 34441401 Free PMC article. Review.
-
Validation of corneal topographic and aberrometric measurements obtained by color light-emitting diode reflection topography in healthy eyes.Graefes Arch Clin Exp Ophthalmol. 2019 Nov;257(11):2437-2447. doi: 10.1007/s00417-019-04453-5. Epub 2019 Sep 3. Graefes Arch Clin Exp Ophthalmol. 2019. PMID: 31482276
References
-
- Ferrer-Blasco T, Montes-Mico R, Peixoto-de-Matos SC, Gonzalez-Meijome JM, Cervino A. Prevalence of corneal astigmatism before cataract surgery. J Cataract Refract Surg. 2009;35(1):70–75. - PubMed
-
- Hoffmann PC, Hutz WW. Analysis of biometry and prevalence data for corneal astigmatism in 23,239 eyes. J Cataract Refract Surg. 2010;36(9):1479–1485. - PubMed
-
- Khan MI, Muhtaseb M. Prevalence of corneal astigmatism in patients having routine cataract surgery at a teaching hospital in the United Kingdom. J Cataract Refract Surg. 2011;37(10):1751–1755. - PubMed
-
- Villegas EA, Alcon E, Artal P. Minimum amount of astigmatism that should be corrected. J Cataract Refract Surg. 2014;40(1):13–19. - PubMed
-
- Laurendeau C, Lafuma A, Berdeaux G. Modelling lifetime cost consequences of toric compared with standard IOLs in cataract surgery of astigmatic patients in four European countries. J Med Econ. 2009;12(3):230–237. - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources