Aneurysm Study of Pipeline in an Observational Registry (ASPIRe)
- PMID: 27610126
- PMCID: PMC4934472
- DOI: 10.1159/000446503
Aneurysm Study of Pipeline in an Observational Registry (ASPIRe)
Abstract
Background and objective: Few prospective studies exist evaluating the safety and efficacy of the Pipeline Embolization Device (PED) in the treatment of intracranial aneurysms. The Aneurysm Study of Pipeline In an observational Registry (ASPIRe) study prospectively analyzed rates of complete aneurysm occlusion and neurologic adverse events following PED treatment of intracranial aneurysms.
Materials and methods: We performed a multicenter study prospectively evaluating patients with unruptured intracranial aneurysms treated with PED. Primary outcomes included (1) spontaneous rupture of the Pipeline-treated aneurysm; (2) spontaneous nonaneurysmal intracranial hemorrhage (ICH); (3) acute ischemic stroke; (4) parent artery stenosis, and (5) permanent cranial neuropathy. Secondary endpoints were (1) treatment success and (2) morbidity and mortality at the 6-month follow-up. Vascular imaging was evaluated at an independent core laboratory.
Results: One hundred and ninety-one patients with 207 treated aneurysms were included in this registry. The mean aneurysm size was 14.5 ± 6.9 mm, and the median imaging follow-up was 7.8 months. Twenty-four aneurysms (11.6%) were small, 162 (78.3%) were large and 21 (10.1%) were giant. The median clinical follow-up time was 6.2 months. The neurological morbidity rate was 6.8% (13/191), and the neurological mortality rate was 1.6% (3/191). The combined neurological morbidity/mortality rate was 6.8% (13/191). The most common adverse events were ischemic stroke (4.7%, 9/191) and spontaneous ICH (3.7%, 7/191). The complete occlusion rate at the last follow-up was 74.8% (77/103).
Conclusions: Our prospective postmarket study confirms that PED treatment of aneurysms in a heterogeneous patient population is safe with low rates of neurological morbidity and mortality. Patients with angiographic follow-up had complete occlusion rates of 75% at 8 months.
Keywords: Flow diverter; Intracranial aneurysm; Neurological morbidity and mortality.
References
-
- Arrese I, Sarabia R, Pintado R, et al. Flow-diverter devices for intracranial aneurysms: systematic review and meta-analysis. Neurosurgery. 2013;73:193–200. - PubMed
-
- Briganti F, Napoli M, Tortora F, et al. Italian multicenter experience with flow-diverter devices for intracranial unruptured aneurysm treatment with periprocedural complications - a retrospective data analysis. Neuroradiology. 2012;54:1145–1152. - PubMed
-
- Brinjikji W, Murad MH, Lanzino G, et al. Endovascular treatment of intracranial aneurysms with flow diverters: a meta-analysis. Stroke. 2013;44:442–447. - PubMed
-
- Yu SC, Kwok CK, Cheng PW, et al. Intracranial aneurysms: midterm outcome of pipeline embolization device - a prospective study in 143 patients with 178 aneurysms. Radiology. 2012;265:893–901. - PubMed
-
- Chan TT, Chan KY, Pang PK, et al. Pipeline embolisation device for wide-necked internal carotid artery aneurysms in a hospital in Hong Kong: preliminary experience. Hong Kong Med J. 2011;17:398–404. - PubMed