. 2017 Jan;50(1-2):73-79.
doi: 10.1016/j.clinbiochem.2016.09.011. Epub 2016 Sep 13.

Measurement of levofloxacin in human plasma samples for a reliable and accessible drug monitoring

Affiliations

Measurement of levofloxacin in human plasma samples for a reliable and accessible drug monitoring

Othoniel Hugo Aragon-Martinez et al. Clin Biochem. 2017 Jan.

Abstract

Objectives: Levofloxacin monitoring is recommended to obtain clinical cure and low incidence of antimicrobial resistance. During the monitoring procedure, levofloxacin should be measured in plasma samples and several assays are reported for this purpose. However, those methods do not have all of the characteristics for an accessible and reliable drug monitoring. For this reason, we develop a method that has all of the essential characteristics for levofloxacin monitoring.

Design and methods: The procedure of validation was done in terms of Food and Drug Administration guidelines. Subsequently, our assay was applied in plasma samples obtained from healthy volunteers with a single oral administration of levofloxacin as well as patients with respiratory diseases under levofloxacin therapy.

Results: Levofloxacin was extracted from samples using only two precipitation steps. Our assay had a rapid run time (5min), adequate sensitivity (0.05μg/ml of lower limit of quantification), and acceptable parameters of validation. Moreover, compound identities were supported using three dimensional spectra and purities were confirmed employing similarity factors (values>900). Variable concentrations of levofloxacin in samples were observed during the application.

Conclusions: Levofloxacin is successfully quantified using our method that shows reliable results, appropriate range, rapid analyses, and cost-effective measurements under a simple and easy technique while all prior methods did not have it all together. Consequently, our method is a valuable tool for routine drug monitoring. Moreover, a complete evaluation of specificity was done for levofloxacin in plasma samples for the first time. Meanwhile, the application data supported the necessity of levofloxacin monitoring.

Keywords: Diode array technology; Drug monitoring; Levofloxacin; Liquid chromatography; Validation.