. 2017 Aug;30(4):277-284.
doi: 10.1080/08941939.2016.1240271. Epub 2016 Oct 26.

Which Suture Material is Optimal for Pancreaticojejunostomy Anastomosis? An In Vitro Study

Affiliations
Free article

Which Suture Material is Optimal for Pancreaticojejunostomy Anastomosis? An In Vitro Study

Kerem Karaman et al. J Invest Surg. 2017 Aug.
Free article

Abstract

Background: Which suture material is optimal for pancreaticojejunostomy (PJ) anastomosis is a matter of debate with contradictory results. The aim of the present in vitro study was to determine the effects of pancreatic juice, bile, and their mixture on different suture materials in terms of breaking strength and disintegration.

Material and methods: Four suture materials, silk, polyglactin 910, polydioxanone, and polypropylene, were tested in pancreatic juice, bile, and their mixture. Determination of breaking strength and disintegration under electron microscope for each suture material was done on days 0, 3, 6, and 10.

Results: The breaking strength of polyglactin 910 and silk was significantly higher than polypropylene and polydioxanone (p < .05). Polyglactin 910 significantly lost its breaking strength with time in pancreatic juice, bile, and their mixture (p < .001). The breaking strength of each type of suture did not significantly alter in pancreatic juice, bile, and their mixture at the baseline measurement and at the end of the experiment (p > .05). No obvious disintegration has been observed under electron microscope in the architecture and appearance of suture materials after days of exposure to pancreatic juice, bile, and their mixture.

Conclusions: None of the suture materials was disintegrated on exposure to pancreatic juice, bile, and their mixture. Polyglactin 910 has the highest breaking strength and significantly loses its strength throughout the experiment but still remains higher than other suture materials. Polypropylene, polydioxanone, and silk showed less variation across the incubation period.

Keywords: bile; breaking strength; disintegration; pancreatic juice; pancreaticojejunostomy anastomosis; suture.