Automatic Oxygen Titration During Walking in Subjects With COPD: A Randomized Crossover Controlled Study
- PMID: 27794080
- DOI: 10.4187/respcare.04406
Automatic Oxygen Titration During Walking in Subjects With COPD: A Randomized Crossover Controlled Study
Abstract
Background: Arterial oxygen desaturation frequently occurs in patients with COPD during daily activities at home. Oxygen flow is usually set at fixed and low rates for ambulatory patients. We evaluated an innovative closed-loop system (FreeO2) that automatically adjusts the oxygen flow to the patient's needs in subjects with COPD during walking followed by recovery time, such as during ambulatory conditions.
Methods: Patients with COPD who exhibited oxygen desaturation on exertion were included in the study. Subjects performed endurance shuttle walk tests followed by 10 min of recovery. The tests were conducted in a random order and in crossover with the 3 following conditions: subjects breathing (1) air at 2 L/min, (2) oxygen at 2 L/min, or (3) FreeO2 (variable oxygen flow). SpO2, pulse rate, PETCO2 , breathing frequency, and oxygen flow were continuously recorded during the 3 conditions. The primary outcome was the percentage of time within the SpO2 target of 92-96%. Secondary outcomes included the endurance shuttle walk test time and distance.
Results: Sixteen subjects with COPD were recruited. The percentage of time with SpO2 in the target range (92-96%) was higher while using the FreeO2, and time with severe oxygen desaturation (SpO2 <88%) was lower with FreeO2 in comparison with constant-flow oxygen and air testing conditions (0.6% vs 23.9% vs 52.2%, P < .001). In comparison with air, walking distance was increased by 35% with oxygen (P = .045) and by 63% with FreeO2 (P < .001). The walking distance was increased by 17% with FreeO2 in comparison with constant oxygen, but the difference was not statistically significant (P = .22).
Conclusions: Automatic titration of oxygen flow during walking to maintain oxygen saturation in a specified range improves oxygenation and may improve exercise tolerance during daily activity, such as walking, in patients with COPD in comparison with room air and fixed oxygen administration. (ClinicalTrials.gov registration: NCT02150434.).
Keywords: COPD; exercise; hyperoxia; hypoxia; oxygen inhalation therapy; technological innovations.
Copyright © 2016 by Daedalus Enterprises.
Similar articles
-
Automated oxygen titration and weaning with FreeO2 in patients with acute exacerbation of COPD: a pilot randomized trial.Int J Chron Obstruct Pulmon Dis. 2016 Aug 24;11:1983-90. doi: 10.2147/COPD.S112820. eCollection 2016. Int J Chron Obstruct Pulmon Dis. 2016. PMID: 27601891 Free PMC article. Clinical Trial.
-
Detecting oxygen desaturation in patients with COPD: incremental versus endurance shuttle walking.Respir Med. 2008 Aug;102(8):1148-52. doi: 10.1016/j.rmed.2008.03.007. Epub 2008 Jun 24. Respir Med. 2008. PMID: 18571914 Clinical Trial.
-
Automated O2 titration improves exercise capacity in patients with hypercapnic chronic obstructive pulmonary disease: a randomised controlled cross-over trial.Thorax. 2019 Mar;74(3):298-301. doi: 10.1136/thoraxjnl-2018-211967. Epub 2018 Aug 30. Thorax. 2019. PMID: 30166425 Clinical Trial.
-
Automatically titrating oxygen system versus constant flow oxygen system during exercise in patients with COPD: a systematic review and meta-analysis.BMC Pulm Med. 2025 Mar 28;25(1):140. doi: 10.1186/s12890-025-03594-0. BMC Pulm Med. 2025. PMID: 40155894 Free PMC article.
-
Exertional desaturation in patients with chronic obstructive pulmonary disease.COPD. 2009 Dec;6(6):478-87. doi: 10.3109/15412550903341497. COPD. 2009. PMID: 19938972 Review.
Cited by
-
Automated oxygen control with O2matic® during admission with exacerbation of COPD.Int J Chron Obstruct Pulmon Dis. 2018 Dec 14;13:3997-4003. doi: 10.2147/COPD.S183762. eCollection 2018. Int J Chron Obstruct Pulmon Dis. 2018. PMID: 30587955 Free PMC article. Clinical Trial.
-
Effect of Automated Oxygen Titration during Walking on Dyspnea and Endurance in Chronic Hypoxemic Patients with COPD: A Randomized Crossover Trial.J Clin Med. 2021 Oct 20;10(21):4820. doi: 10.3390/jcm10214820. J Clin Med. 2021. PMID: 34768338 Free PMC article.
-
Automated versus manual oxygen titration in COPD exacerbation: machine or hands, this is the question.Int J Chron Obstruct Pulmon Dis. 2017 Apr 3;12:1057-1060. doi: 10.2147/COPD.S130107. eCollection 2017. Int J Chron Obstruct Pulmon Dis. 2017. PMID: 28435238 Free PMC article. No abstract available.
-
Supplemental oxygen for symptomatic relief in people with serious respiratory illness: a systematic review and meta-analysis.Eur Respir Rev. 2025 Mar 5;34(175):240025. doi: 10.1183/16000617.0025-2024. Print 2025 Jan. Eur Respir Rev. 2025. PMID: 40044186 Free PMC article.
-
Automated oxygen administration versus manual control in acute cardiovascular care: a randomised controlled trial.Heart. 2024 Dec 13;111(1):27-34. doi: 10.1136/heartjnl-2024-324488. Heart. 2024. PMID: 39486892 Free PMC article. Clinical Trial.
Publication types
MeSH terms
Associated data
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical