Aims: We present a minimum 20-year follow-up study of 382 cemented Exeter Universal total hip arthroplasties (350 patients) operated on at a mean age of 66.3 years (17 to 94).
Patients and methods: All patients received the same design of femoral component, regardless of the original diagnosis. Previous surgery had been undertaken for 33 hips (8.6%). During the study period 218 patients with 236 hips (62%) died, 42 hips (11%) were revised and 110 hips (29%) in 96 patients were available for review. The acetabular components were varied and some designs are now obsolete, however they were all cemented.
Results: With an endpoint of revision for aseptic loosening or lysis, survivorship of the stem at 22.8 years was 99.0% (95% confidence interval (CI) 97.0 to 100). One stem was revised 21 years post-operatively in a patient with Gaucher's disease and proximal femoral osteolysis. Survivorship with aseptic loosening or lysis of the acetabular component or stem as the endpoint at 22.8 years was 89.3% (95% CI 84.8 to 93.8). With an endpoint of revision for any reason, overall survivorship was 82.9% (95% CI 77.4 to 88.4) at 22.8 years. Radiological review showed excellent preservation of bone stock at 20 to 25 years, and no impending failures of the stem.
Conclusion: The Exeter femoral stem continues to perform well beyond 20 years. Cite this article: Bone Joint J 2016;98-B:1441-9.
Keywords: Cement; Exeter hip; Long-term follow-up; Primary total hip arthroplasty; Survival.
©2016 The British Editorial Society of Bone & Joint Surgery.
Similar articles
-
The Exeter Contemporary flanged cemented acetabular component in primary total hip arthroplasty.Bone Joint J. 2016 Mar;98-B(3):307-12. doi: 10.1302/0301-620X.98B3.35901. Bone Joint J. 2016. PMID: 26920954
-
Exeter Universal cemented femoral component.Bone Joint J. 2020 Oct;102-B(10):1319-1323. doi: 10.1302/0301-620X.102B10.BJJ-2019-1454.R1. Bone Joint J. 2020. PMID: 32993348
-
The Exeter V40 cemented femoral component at a minimum 10-year follow-up: the first 540 cases.Bone Joint J. 2018 Aug;100-B(8):1002-1009. doi: 10.1302/0301-620X.100B8.BJJ-2017-1535.R1. Bone Joint J. 2018. PMID: 30062940
-
Revision total hip arthroplasty: the femoral side using cemented implants.Int Orthop. 2011 Feb;35(2):267-73. doi: 10.1007/s00264-010-1167-5. Epub 2010 Dec 17. Int Orthop. 2011. PMID: 21165618 Free PMC article. Review.
-
Femoral stem wear in cemented total hip replacement.Proc Inst Mech Eng H. 2008 Jul;222(5):583-92. doi: 10.1243/09544119JEIM346. Proc Inst Mech Eng H. 2008. PMID: 18756677 Review.
Cited by
-
Periprosthetic femoral fractures around the original cemented polished triple-tapered C-stem femoral implant: a consecutive series of 500 primary total hip arthroplasties with an average follow-up of 15 years.Arch Orthop Trauma Surg. 2023 Jul;143(7):4511-4518. doi: 10.1007/s00402-022-04712-x. Epub 2022 Nov 29. Arch Orthop Trauma Surg. 2023. PMID: 36447057 Free PMC article.
-
Fourteen-year experience with short cemented stems in total hip replacement.Int Orthop. 2019 Jan;43(1):55-61. doi: 10.1007/s00264-018-4205-3. Epub 2018 Nov 8. Int Orthop. 2019. PMID: 30411248
-
Experience of an anatomic femoral stem in a UK orthopaedic centre beyond 20 years of follow-up.Eur J Orthop Surg Traumatol. 2024 May;34(4):2155-2162. doi: 10.1007/s00590-024-03901-1. Epub 2024 Apr 3. Eur J Orthop Surg Traumatol. 2024. PMID: 38565783 Free PMC article.
-
The effects of patient characteristics and stem alignment on distal femoral cortical hypertrophy after cemented polished tapered stem implantation.Eur J Orthop Surg Traumatol. 2020 May;30(4):559-567. doi: 10.1007/s00590-019-02605-1. Epub 2019 Dec 19. Eur J Orthop Surg Traumatol. 2020. PMID: 31853636
-
Peri-prosthetic bone remodeling and change in bone mineral density in the femur after cemented polished tapered stem implantation.Eur J Orthop Surg Traumatol. 2019 Jul;29(5):1061-1067. doi: 10.1007/s00590-019-02414-6. Epub 2019 Mar 8. Eur J Orthop Surg Traumatol. 2019. PMID: 30848380
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical