Measurement of endotracheal tube cuff pressure: Instrumental versus conventional method

Affiliations

Measurement of endotracheal tube cuff pressure: Instrumental versus conventional method

Mueen Ullah Khan et al. Saudi J Anaesth. 2016 Oct-Dec.

Abstract

Objective: To evaluate the conventional practice of endotracheal tube (ETT) cuff inflation and pressure measurement as compared to the instrumental method.

Study design: Prospective observational study.

Place and duration of study: Department of Anaesthesia, King Saud University Hospital, Riyadh, Saudi Arabia (June 2014-July 2014).

Methods: A total of 100 adult patients were observed according to the syringe size used Group-1 (10 ml) and Group-2 (20 ml) for ETT cuff inflation in general anesthesia. Patients with anticipated difficult intubation, risk for aspiration, known anatomical laryngotracheal abnormalities, and emergency cases were excluded. Trachea was intubated with size 8 or 8.5 mm and 7.0 or 7.5 mm ETT in male and female patients respectively. The ETT cuff was inflated with air by one of the anesthesia technician. Cuff pressures were measured using aneroid manometer. ETT cuff pressure of 20-30 cm of water was considered as standard.

Results: In 69% of the patients, the cuff pressure measurements were above the standard. Age (P = 0.806), weight (P = 0.527), height (P = 0.850), and gender (P = 1.00) were comparable in both groups. The mean cuff pressure in Group-1 and Group-2 was 32.52 ± 6.39 and 38.90 ± 6.60 cm of water (P = 0.001). The cuff inflation with 20 ml syringe resulted in higher cuff pressure as compared to 10cc syringe 37.73 ± 4.23 versus 40.74 ± 5.01 (86% vs. 52%, P = 0.013).

Conclusion: The conventional method for ETT cuff inflation and pressure measuring is unreliable. As a routine instrumental cuff pressure, monitoring is suggested.

Keywords: Cuff pressure; endotracheal tube; measurement.

PubMed Disclaimer

References

    1. Sengupta P, Sessler DI, Maglinger P, Wells S, Vogt A, Durrani J, et al. Endotracheal tube cuff pressure in three hospitals, and the volume required to produce an appropriate cuff pressure. BMC Anesthesiol. 2004;4:8. - PMC - PubMed
    1. Sole ML, Su X, Talbert S, Penoyer DA, Kalita S, Jimenez E, et al. Evaluation of an intervention to maintain endotracheal tube cuff pressure within therapeutic range. Am J Crit Care. 2011;20:109–17. - PMC - PubMed
    1. Nseir S, Brisson H, Marquette CH, Chaud P, Di Pompeo C, Diarra M, et al. Variations in endotracheal cuff pressure in intubated critically ill patients: Prevalence and risk factors. Eur J Anaesthesiol. 2009;26:229–34. - PubMed
    1. Blot SI, Poelaert J, Kollef M. How to avoid microaspiration? A key element for the prevention of ventilator-associated pneumonia in intubated ICU patients. BMC Infect Dis. 2014;14:119. - PMC - PubMed
    1. American Thoracic Society; Infectious Diseases Society of America. Guidelines for the management of adults with hospital-acquired, ventilator-associated, and healthcare-associated pneumonia. Am J Respir Crit Care Med. 2005;171:388–416. - PubMed