Assessing agreement of blood pressure-measuring devices in Tibetan areas of China: a systematic review
- PMID: 27843497
- PMCID: PMC5093356
- DOI: 10.1136/heartasia-2016-010798
Assessing agreement of blood pressure-measuring devices in Tibetan areas of China: a systematic review
Abstract
Background: The validity of blood pressure (BP)-measuring tools at very high altitudes is uncertain. Therefore, the objective of this review was to examine the degree of agreement of BP-measuring devices in Tibet.
Methods: We conducted electronic searches in Medline, Embase, Cinahl, Cochrane Library, Global Health Library and the ISI Web of Science. Randomised and observational studies were considered for inclusion. The methodological characteristics of included studies were assessed using the Quality Assessment of Diagnostic Accuracy Studies-2 criteria. Our primary outcome was the difference in mean BP measurements between the new device and the gold standard.
Results: We identified three eligible studies, out of which two with a total of 162 participants were included. The studies differed in their methodology. One study reported significantly higher systolic blood pressure (SBP) measurement with electronic sphygmomanometer (Omron) compared with mercury sphygmomanometer (mean difference 5.8±4.7 mm Hg; p<0.001), with no significant difference in diastolic blood pressure (DBP) measurement (0.4±3.9 mm Hg; p=0.23). The second study reported mean differences of 1.0±5.9 mm Hg and -3.1±4.6 mm Hg for SBP and DBP, respectively.
Conclusion: The limited evidence from published studies suggests that automated (Omron) BP monitors show a high degree of agreement for DBP when compared against mercury sphygmomanometer at high altitudes. However, the degree of such agreement for SBP is not consistent. Few studies assessing the validity of automated BP monitors at high altitudes have been conducted, and they differ in design and methodology. Further research assessing the suitability of BP-measuring instruments at high altitudes is therefore warranted.
Keywords: EBM; HYPERTENSION.
Conflict of interest statement
None
Figures


Similar articles
-
Validation of the Omron M6 (HEM-7001-E) upper-arm blood pressure measuring device according to the International Protocol in adults and obese adults.Blood Press Monit. 2007 Aug;12(4):219-25. doi: 10.1097/MBP.0b013e3280f813d0. Blood Press Monit. 2007. PMID: 17625394
-
Can an automatic oscillometric device replace a mercury sphygmomanometer on blood pressure measurement? a systematic review and meta-analysis.Blood Press Monit. 2019 Dec;24(6):265-276. doi: 10.1097/MBP.0000000000000412. Blood Press Monit. 2019. PMID: 31658107
-
Blood pressure randomized methodology study comparing automatic oscillometric and mercury sphygmomanometer devices: National Health and Nutrition Examination Survey, 2009-2010.Natl Health Stat Report. 2012 Oct 5;(59):1-15. Natl Health Stat Report. 2012. PMID: 24984529 Clinical Trial.
-
A comparative study of an ambulatory blood pressure measuring device and a wrist blood pressure monitor with a position sensor versus a mercury sphygmomanometer.Eur J Intern Med. 2007 Mar;18(2):118-23. doi: 10.1016/j.ejim.2006.09.018. Eur J Intern Med. 2007. PMID: 17338963
-
Assessing the validity of oscillometric device for blood pressure measurement in a large population-based epidemiologic study.J Am Soc Hypertens. 2017 Nov;11(11):730-736.e4. doi: 10.1016/j.jash.2017.09.004. Epub 2017 Sep 21. J Am Soc Hypertens. 2017. PMID: 29032943
Cited by
-
Applicability of electronic sphygmomanometer use in high-altitude areas according to the AAMI/ESH/ISO collaboration statement.Front Cardiovasc Med. 2024 Jan 8;10:1257444. doi: 10.3389/fcvm.2023.1257444. eCollection 2023. Front Cardiovasc Med. 2024. PMID: 38259316 Free PMC article.
-
Validation of Omron HBP-1100-E Professional Blood Pressure Measuring Device According to the American Association for the Advancement of Medical Instrumentation Protocol: The PERSIAN Guilan Cohort Study (PGCS).Med Devices (Auckl). 2020 Sep 3;13:231-236. doi: 10.2147/MDER.S253638. eCollection 2020. Med Devices (Auckl). 2020. PMID: 32943949 Free PMC article.
References
-
- Lewington S, Clarke R, Qizilbash N et al. . Age specific relevance of usual blood pressure to vascular mortality: a meta-analysis of individual data for 1 million adults in 61 prospective studies. Lancet 2002;360:1903–13. doi:10.1016/S0140-6736(02)11911-8 - DOI - PubMed
-
- Kearney PM, Whelton M, Reynolds K et al. . Global burden of hypertension: analysis of worldwide data. Lancet 2005;365:217–23. doi:10.1016/S0140-6736(05)17741-1 - DOI - PubMed
-
- World Health Organization. A global brief on hypertension (document number WHO/DCO/WHD/2013.2) World Health Organization, Geneva (2013). http://ish-world.com/downloads/pdf/global_brief_hypertension.pdf (accessed 20 May 2015).
-
- World Health Organization. Global status reports on non-communicable diseases 2014. http://apps.who.int/iris/bitstream/10665/148114/1/9789241564854_eng.pdf?... (accessed 20 May 2015).
-
- World Health Organization. Global action plan for the prevention and control of non-communicable diseases 2013-2020. WHO, 2013. (accessed 20 May 2015). http://apps.who.int/iris/bitstream/10665/94384/1/9789241506236_eng.pdf?ua=1