Effects of hip implant modular neck material and assembly method on fatigue life and distraction force
- PMID: 27862223
- DOI: 10.1002/jor.23481
Effects of hip implant modular neck material and assembly method on fatigue life and distraction force
Abstract
Hip implant neck fractures and adverse tissue reactions associated with fretting-corrosion damage at modular interfaces are a major source of concern. Therefore, there is an urgent clinical need to develop accurate in vitro test procedures to better understand, predict and prevent in vivo implant failures. This study aimed to simulate in vivo fatigue fracture and distraction of modular necks in an in vitro setting, and to assess the effects of neck material (Ti6Al4V vs. CoCrMo) and assembly method (hand vs. impact) on the fatigue life and distraction of the necks. Fatigue tests were performed on the cementless PROFEMUR® Total Hip Modular Neck System under two different loads and number of cycles: 2.3 kN for 5 million cycles, and 7.0 kN for 1.3 million cycles. The developed in vitro simulation setup successfully reproduced in vivo modular neck fracture mode and location. Neck failure occurred at the neck-stem taper and the fracture ran from the distal lateral neck surface to the proximal medial entry point of the neck into the stem. None of the necks failed under the 2.3 kN load. However, all hand-assembled Ti6Al4V necks failed under the 7.0 kN load. In contrast, none of the hand-assembled CoCrMo necks and impact-assembled necks (Ti6Al4V or CoCrMo) failed under this higher load. In conclusion, Ti6Al4V necks were more susceptible to fatigue failure than CoCrMo necks. In addition, impact assembly substantially improved the fatigue life of Ti6Al4V necks and also led to overall higher distraction forces for both neck materials. Overall, this study shows that the material and assembly method can affect the fatigue strength of modular necks. Finally, improper implant assembly during surgery may result in diminished modular neck survivability and increased failure rates. © 2016 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 35:2023-2030, 2017.
Keywords: assembly; fracture; hip implant; material; modular neck.
© 2016 Orthopaedic Research Society. Published by Wiley Periodicals, Inc.
Similar articles
-
In vitro assessment of strength, fatigue durability, and disassembly of Ti6Al4V and CoCrMo necks in modular total hip replacements.J Biomed Mater Res B Appl Biomater. 2011 Apr;97(1):132-8. doi: 10.1002/jbm.b.31794. Epub 2011 Feb 2. J Biomed Mater Res B Appl Biomater. 2011. PMID: 21290591
-
Effect of impact assembly on the interface deformation and fretting corrosion of modular hip tapers: An in vitro study.J Orthop Res. 2018 Jan;36(1):405-416. doi: 10.1002/jor.23601. Epub 2017 Jun 26. J Orthop Res. 2018. PMID: 28485507
-
Assembly force and taper angle difference influence the relative motion at the stem-neck interface of bi-modular hip prostheses.Proc Inst Mech Eng H. 2016 Jul;230(7):690-9. doi: 10.1177/0954411916648717. Epub 2016 May 10. Proc Inst Mech Eng H. 2016. PMID: 27170535
-
Non-modular tapered fluted titanium stems in hip revision surgery: gaining attention.Bone Joint J. 2014 Nov;96-B(11 Supple A):56-9. doi: 10.1302/0301-620X.96B11.34302. Bone Joint J. 2014. PMID: 25381409 Review.
-
Modular taper junction corrosion and failure: how to approach a recalled total hip arthroplasty implant.J Arthroplasty. 2014 Jan;29(1):1-6. doi: 10.1016/j.arth.2013.08.026. Epub 2013 Sep 30. J Arthroplasty. 2014. PMID: 24090661 Review.
Cited by
-
Microstructure and electrochemical behavior of contemporary Ti6Al4V implant alloys.J Bio Tribocorros. 2022 Mar;8(1):26. doi: 10.1007/s40735-021-00623-3. Epub 2021 Dec 24. J Bio Tribocorros. 2022. PMID: 35911172 Free PMC article.
-
What Factors Are Associated With Neck Fracture in One Commonly Used Bimodular THA Design? A Multicenter, Nationwide Study in Slovenia.Clin Orthop Relat Res. 2019 Jun;477(6):1324-1332. doi: 10.1097/CORR.0000000000000646. Clin Orthop Relat Res. 2019. PMID: 31136429 Free PMC article.
-
Additively manufactured metallic biomaterials.Bioact Mater. 2021 Dec 30;15:214-249. doi: 10.1016/j.bioactmat.2021.12.027. eCollection 2022 Sep. Bioact Mater. 2021. PMID: 35386359 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous