Single-incision laparoscopic surgery increases the risk of unintentional thermal injury from the monopolar "Bovie" instrument in comparison with traditional laparoscopy
- PMID: 27864716
- DOI: 10.1007/s00464-016-5339-2
Single-incision laparoscopic surgery increases the risk of unintentional thermal injury from the monopolar "Bovie" instrument in comparison with traditional laparoscopy
Abstract
Background: Single-incision laparoscopic surgery (SILS) places multiple instruments in close, parallel proximity, an orientation that may have implications in the production of stray current from the monopolar "Bovie" instrument. The purpose of this study was to compare the energy transferred during SILS compared to traditional four-port laparoscopic surgery (TRD).
Method: In a laparoscopic simulator, instruments were inserted via SILS or TRD setup. The monopolar generator delivered energy to a laparoscopic L-hook instrument for 5-s activations on 30-Watts coag mode. The primary outcome (stray current) was quantified by measuring the heat of liver tissue held adjacent to the non-electrically active 10-mm telescope tip and Maryland grasper in both the SILS and TRD setups. To control for the potential confounder of stray energy coupling via wires outside the surgical field, the camera cord and active electrode wires were oriented parallel or completely separated.
Results: SILS and TRD setups create similar amounts of stray current as measured by increased tissue temperature at the non-electrically active telescope tip (41 ± 12 vs. 39 ± 10 °C; p = 0.71). Stray current was greater in SILS compared to TRD at the tip of the non-electrically active Maryland forceps (38 ± 9 vs. 20 ± 10 °C; p < 0.01). Separation of the active electrode and camera cords did not change the amount of stray energy in the SILS orientation for either telescope (39 ± 10 °C bundled vs. 36 ± 10 °C separated; p = 0.40) or grasper (38 ± 9 °C bundled vs. 34 ± 11 °C separated; p = 0.19) but did in the TRD orientation (41 ± 12 bundled vs. 24 ± 10 separated; p < 0.01). When SILS was compared to TRD with the cords separated, SILS increased stray energy at both the telescope tip and grasper tip (36 ± 10 vs. 24 ± 10 °C; p < 0.01 and 34 ± 11 vs. 17 ± 8 °C; p < 0.01).
Conclusion: SILS increases stray energy transfer nearly twice as much as TRD with the use of the monopolar instrument. Strategies to mitigate the amount of stray energy in the TRD setup such as separation of the active electrode and camera cords are not effective in the SILS setup. These practical findings should enhance surgeons using the SILS approach of increased stray energy that could result in injury.
Keywords: Antenna coupling; Capacitive coupling; Electrosurgery; Patient safety; Radiofrequency energy.
Similar articles
-
Hand-to-hand coupling and strategies to minimize unintentional energy transfer during laparoscopic surgery.J Surg Res. 2017 Nov;219:103-107. doi: 10.1016/j.jss.2017.05.091. Epub 2017 Jun 28. J Surg Res. 2017. PMID: 29078867 Clinical Trial.
-
Unintended stray energy from monopolar instruments: beware the dispersive electrode cord.Surg Endosc. 2016 Apr;30(4):1333-6. doi: 10.1007/s00464-015-4388-2. Epub 2015 Jul 15. Surg Endosc. 2016. PMID: 26173544
-
Radiofrequency energy antenna coupling to common laparoscopic instruments: practical implications.Surg Endosc. 2012 Nov;26(11):3053-7. doi: 10.1007/s00464-012-2312-6. Epub 2012 May 12. Surg Endosc. 2012. PMID: 22580879
-
Active electrode monitoring. How to prevent unintentional thermal injury associated with monopolar electrosurgery at laparoscopy.Surg Endosc. 1998 Aug;12(8):1009-12. doi: 10.1007/s004649900769. Surg Endosc. 1998. PMID: 9685531 Review.
-
Electrosurgery: principles and safety issues.Clin Obstet Gynecol. 1995 Sep;38(3):610-21. doi: 10.1097/00003081-199509000-00021. Clin Obstet Gynecol. 1995. PMID: 8612371 Review.
Cited by
-
Comparing the safety and effectiveness of dedicated radiofrequency transseptal wires to electrified metal guidewires.J Cardiovasc Electrophysiol. 2022 Mar;33(3):371-379. doi: 10.1111/jce.15341. Epub 2022 Jan 11. J Cardiovasc Electrophysiol. 2022. PMID: 34978365 Free PMC article.
-
Stray energy injury during robotic versus laparoscopic inguinal hernia repair: a randomized controlled trial.Surg Endosc. 2023 Nov;37(11):8771-8777. doi: 10.1007/s00464-023-10331-3. Epub 2023 Aug 14. Surg Endosc. 2023. PMID: 37580577 Clinical Trial.
-
Monopolar stray energy in robotic surgery.Surg Endosc. 2021 May;35(5):2084-2090. doi: 10.1007/s00464-020-07605-5. Epub 2020 May 8. Surg Endosc. 2021. PMID: 32385708
-
The SAGES Fundamental Use of Surgical Energy program (FUSE): history, development, and purpose.Surg Endosc. 2018 Jun;32(6):2583-2602. doi: 10.1007/s00464-017-5933-y. Epub 2017 Dec 7. Surg Endosc. 2018. PMID: 29218661
-
Comparison of the effects of an ERAS program and a single-port laparoscopic surgery on postoperative outcomes of colon cancer patients.Sci Rep. 2019 Aug 19;9(1):11998. doi: 10.1038/s41598-019-48526-1. Sci Rep. 2019. PMID: 31427651 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical