. 2017 Sep;31(5):e22099.
doi: 10.1002/jcla.22099. Epub 2016 Nov 21.

National continuous surveys on internal quality control for HbA1c in 306 clinical laboratories of China from 2012 to 2016: Continual improvement

Affiliations

National continuous surveys on internal quality control for HbA1c in 306 clinical laboratories of China from 2012 to 2016: Continual improvement

Tingting Li et al. J Clin Lab Anal. 2017 Sep.

Abstract

Background: This study aimed to evaluate whether the quality performance of clinical laboratories in China has been greatly improved and whether Internal Quality Control (IQC) practice of HbA1c has also been changed since National Center for Clinical Laboratories (NCCL) of China organized laboratories to report IQC data for HbA1c in 2012.

Methods: Internal Quality Control information of 306 External Quality Assessment (EQA) participant laboratories which kept reporting IQC data in February from 2012 to 2016 were collected by Web-based EQA system. Then percentages of laboratories meeting four different imprecision specifications for current coefficient of variations (CVs) of HbA1c measurements were calculated. Finally, we comprehensively analyzed analytical systems and IQC practice of HbA1c measurements.

Results: The current CVs of HbA1c tests have decreased significantly from 2012 to 2016. And percentages of laboratories meeting four imprecision specifications for CVs all showed the increasing tendency year by year. As for analytical system, 52.1% (159/306) laboratories changed their systems with the change in principle of assay. And many laboratories began to use cation exchange high-performance liquid chromatography (CE-HPLC) instead of Immunoturbidimetry, because CE-HPLC owed a lower intra-laboratory CVs. The data of IQC practice, such as IQC rules and frequency, also showed significant variability among years with overall tendency of meeting requirements.

Conclusion: The imprecision performance of HbA1c tests has been improved in these 5 years with the change in IQC practice, but it is still disappointing in China. Therefore, laboratories should actively find existing problems and take action to promote performance of HbA1c measurements.

Keywords: HbA1c; clinical laboratory; coefficient of variation; diabetes; diagnosis; imprecision; internal quality control; quality specification.

PubMed Disclaimer

References

    1. American Diabetes Association . Report of the expert committee on the diagnosis and classification of diabetes mellitus. Diabetes Care. 1997;20:1183–1197. - PubMed
    1. IDF . Diabetes Atlas, 3rd edn Brussels: IDF; 2008.
    1. American Diabetes Association . Diagnosis and classification of diabetes mellitus. Diabetes Care. 2010;33:S62–S69. - PMC - PubMed
    1. Zeng R, Wang W, Zhao H, Fei Y, Wang Z. National survey on internal quality control for HbA1c analytical instruments in 331 Hospital Laboratories of China. Clin Lab. 2015;61:453–460. - PubMed
    1. International Organization for Standardization . Medical laboratories ‐Requirements for quality and competence. ISO 15189:2012. - PubMed

Substances

LinkOut - more resources