Clinical Evaluation and Outcomes of Digital Chest Drainage after Lung Resection
- PMID: 27885215
- PMCID: PMC5183980
- DOI: 10.5761/atcs.oa.16-00179
Clinical Evaluation and Outcomes of Digital Chest Drainage after Lung Resection
Abstract
Background: Analog chest drainage systems (ACS) are generally used to monitor postoperative alveolar air leakage (PAL) after lung resection. An electronic digital chest drainage system (DCS) has recently been developed that reportedly has several advantages over the traditional ACS. Here, we report a single institution's experience of PAL management with the DCS. We also sought to establish whether DCS had superior clinical benefits and outcomes compared with ACS.
Methods: We enrolled 112 consecutive patients who underwent lung resection and were subsequently managed with DCS. We compared PAL rate, duration of chest drainage, and the incidence of complications with a group of 121 consecutive patients previously managed with ACS after lung resection, using propensity score matching.
Results: Mean maximum and minimum PAL rates during DCS chest drainage were 48.9 ml/min (range: 2.0-868.6 ml/min) and 0.1 ml/min (0.0-1.2 ml/min), respectively. Mean PAL rate at DCS removal was 1.3 ml/min (0.0-10.0 ml/min). After propensity score matching, mean duration of chest drainage was significantly shorter with DCS than ACS (2.7 days, range: 1-9 days, compared with 3.7 days, range: 1-21 days, respectively; P = 0.031).
Conclusions: Managing PAL with DCS after pulmonary resection appears to reduce the duration of chest drainage.
References
-
- Cerfolio RJ, Bryant AS. The benefits of continuous and digital air leak assessment after elective pulmonary resection: a prospective study. Ann Thorac Surg 2008; 86: 396-401. - PubMed
-
- Varela G, Jimenez MF, Novoa NM, et al. Postoperative chest tube management: measuring air leak using an electronic device decreases variability in the clinical practice. Eur J Cardiothorac Surg 2009; 35: 28-31. - PubMed
-
- Anegg U, Lindenmann J, Matzi V, et al. AIRFIX: the first digital postoperative chest tube airflowmetry—a novel method to quantify air leakage after lung resection. Eur J Cardiothorac Surg 2006; 29: 867-72. - PubMed
-
- Brunelli A, Cassivi SD, Salati M, et al. Digital measurements of air leak flow and intrapleural pressures in the immediate postoperative period predict risk of prolonged air leak after pulmonary lobectomy. Eur J Cardiothorac Surg 2011; 39: 584-8. - PubMed
-
- Dernevik L, Belboul A, Rådberg G. Initial experience with the world’s first digital drainage system. The benefits of recording air leaks with graphic representation. Eur J Cardiothorac Surg 2007; 31: 209-13. - PubMed