Arterial pulsatility under phasic left ventricular assist device support
- PMID: 27885993
- DOI: 10.3233/BME-161599
Arterial pulsatility under phasic left ventricular assist device support
Abstract
The aim of this study is to understand whether the phasic Continuous Flow Left Ventricular Assist Device (CF-LVAD) support would increase the arterial pulsatility. A Micromed DeBakey CF-LVAD was used to apply phasic support in an ex-vivo experimental platform. CF-LVAD was operated over a cardiac cycle by phase-shifting the pulsatile pump control with respect to the heart cycle, in 0.05 s increments in each experiment. The pump flow rate was selected as the control variable and a reference model was used to operate the CF-LVAD at a pulsatile speed. Arterial pulse pressure was the highest (9 mmHg) when the peak pump flow is applied at the peak systole under varying speed CF-LVAD support over a cardiac cycle while it was the lowest (2 mmHg) when the peak pump flow was applied in the diastolic phase. The mean arterial pressure and mean CF-LVAD output were the same in each experiment while arterial pulse pressure and pulsatility index varied depending on the phase of reference pump flow rate signal. CF-LVAD speed should be synchronized considering the timing of peak systole over a cardiac cycle to increase the arterial pulsatility. Moreover, it is possible to decrease the arterial pulsatility under counter-pulsating CF-LVAD support.
Keywords: CF-LVAD; phasic CF-LVAD support; pulsatility; varying speed.
Similar articles
-
Arterial pulsatility improvement in a feedback-controlled continuous flow left ventricular assist device: an ex-vivo experimental study.Med Eng Phys. 2014 Oct;36(10):1288-95. doi: 10.1016/j.medengphy.2014.07.005. Epub 2014 Jul 25. Med Eng Phys. 2014. PMID: 25066581
-
Improving arterial pulsatility by feedback control of a continuous flow left ventricular assist device via in silico modeling.Int J Artif Organs. 2014 Oct;37(10):773-85. doi: 10.5301/ijao.5000328. Epub 2014 May 13. Int J Artif Organs. 2014. PMID: 24970558
-
Enhancement of Arterial Pressure Pulsatility by Controlling Continuous-Flow Left Ventricular Assist Device Flow Rate in Mock Circulatory System.J Med Biol Eng. 2016;36:308-315. doi: 10.1007/s40846-016-0140-1. Epub 2016 Jun 25. J Med Biol Eng. 2016. PMID: 27441034 Free PMC article.
-
Clinical Implications of Physiologic Flow Adjustment in Continuous-Flow Left Ventricular Assist Devices.ASAIO J. 2017 May/Jun;63(3):241-250. doi: 10.1097/MAT.0000000000000477. ASAIO J. 2017. PMID: 28459742 Review.
-
The Unique Blood Pressures and Pulsatility of LVAD Patients: Current Challenges and Future Opportunities.Curr Hypertens Rep. 2017 Oct 18;19(10):85. doi: 10.1007/s11906-017-0782-6. Curr Hypertens Rep. 2017. PMID: 29043581 Free PMC article. Review.
Cited by
-
Design and Analysis of a Polymeric Left Ventricular Simulator via Computational Modelling.Biomimetics (Basel). 2024 Apr 28;9(5):269. doi: 10.3390/biomimetics9050269. Biomimetics (Basel). 2024. PMID: 38786479 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources